User guide

2.3 Attaching the accessories
754 Dialysis Unit
11
2
Connect new feedthrough
Using a wrench or a spanner, unscrew the nut on the interior
side of feedthrough 2727 and remove the feedthrough.
Insert new 6.2756.000 tubing feedthrough in the hole and
retighten the nut.
3
Attach tubing to valve A
Screw 6.1825.210 PEEK sample loop (20 µL) to connections
"3" and "6" of valve A 1818 using the enclosed 6.2744.010 PEEK
compression fittings.
Screw PEEK capillary 1919 (6.1831.060; L =1 m) to connection
"2" of valve A 1818 using a 6.2744.010 PEEK compression fit-
ting. Pull the other end of the tubing to the back through the
rear panel opening 4141.
Screw PEEK capillary 2020 (6.1831.040; L =15 cm) to connec-
tion "1" of valve A 18 18 using a 6.2744.010 PEEK compression
fitting.
4
Attach tubing to valve B
Screw the PEEK capillary 2020 (6.1831.040; L =15 cm) at-
tached to connection "1" of valve A 1818 to connection "3" of
valve B 22 22 using a 6.2744.010 PEEK compression fitting.
Screw PEEK capillary 2121 (6.1831.040; L =15 cm) to connec-
tions "1" and "2" of valve B 2222 using two 6.2744.010 PEEK
compression fittings.
Screw PEEK capillary 2525 (6.1831.050; L =40 cm) to connec-
tion "4" of valve B 2222 using a 6.2744.010 PEEK compression
fitting and pull the other end to the front through feedthrough
2727.
Screw PEEK capillary 2424 (6.1831.050; L =40 cm) to connec-
tion "5" of valve B 2222 using a 6.2744.010 PEEK compression
fitting and pull the other end to the front through feedthrough
2727.
Screw PEEK capillary 2323 (6.1831.050; L =40 cm) to connec-
tion "6" of valve B 2222 using a 6.2744.010 PEEK compression
fitting and pull the other end to the front through feedthrough
2828.
5
Connecting to the 753 Suppressor Module (Option)
Insert the suppressor block in the 733 IC Separation Center
as per the 753 Instructions for Use.
Connect tubing to the IC column, the detector block and the
waste in the 733 IC Separation Center as per the 753 Instruc-
tions for Use.
Pull the two inlet capillaries for H
2
SO
4
and H
2
O to the front
through feedthrough 2727 and connect to the 753 Suppressor
Module pump tubings.