User Manual
2.4 Connection to the 732/733 IC system
753 Suppressor Module / 8.753.1003 Instructions for Use
17
8
Fix connection suppressor – detector block
• Insert one of the column holders 8080 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rail 7979 and fasten coupling 9292 in
the column holder.
9
Attach pump tubings
• Loosen both tubing cartridges 44 from the holding clamp 99 by
pressing down snap-action lever 1010 and remove from
mounting pin 1212 on 753 Suppressor Module (see Fig. 1).
• Press contact pressure lever 55 on both tubing cartridges
down as far as it will go.
• Insert a length of pump tubing 3333 (6.1826.050) into each of
the tubing cartridges as shown in Fig. 7. The white-yellow
stopper 3434 must click into the corresponding holder on the
left-hand side of the tubing cartridge.
• Place the tubing cartridges on mounting pin 1212 and press
down on the right-hand side until snap-action lever 1010 clicks
into position on holding clamp 99. Take care that no kinks are
formed in the pump tubing.
10
Suppressor connection 2: H
2
SO
4
• Pull the inlet capillary 2525 marked with “H
2
SO
4
“ at suppressor
connection 2222 (see Fig. 6) through one of the rear panel
openings 4141 or 4343 of the 733 IC Separation Center.
• Mount a PEEK compression fitting 3131 (6.2744.010) on the
end of inlet capillary 2525 and screw this on to coupling 3232
(6.2744.030) (see Fig. 7).
• Push coupling 3232 on to the outlet end of the first pump tubing
3333 (see Fig. 7).
• Mount suction tubing: cut a piece of PTFE Tubing 3030
(6.1803.020) to the desired length.
• Mount a PEEK compression fitting 3131 (6.2744.010) on one
end of PTFE tubing 3030 and screw this on to coupling 3232 (see
Fig. 7).
• Push coupling 3232 on to the inlet end of the first pump tubing
3333 (see Fig. 7).
• Immerse the other end of the suction tubing in a vessel
containing regeneration solution (normally 20 mmol/L H
2
SO
4
)
and fix in place.
• Pull outlet capillary 2626 marked with "Waste" at suppressor
connection 2222 (see Fig. 6) through one of the rear panel
openings 4141 or 4343 of the 733 IC Separation Center.
• Lead outlet capillary 2626 to a sufficiently large waste container
and fix in place.