User Manual

2 Installation
753 Suppressor Module / 8.753.1003 Instructions for Use
16
1
Insert suppressor block
733.0010 IC Separation Center (1-channel system):
Place suppressor block 2121 to the right-hand side of detector
block 7777 (see Fig. 14 of 732/733 Manual) on the floor of the
inner chamber in such a way that suppressor connection 2222
points in the direction of the door.
733.0020 IC Separation Center (two-channel system):
Place suppressor block 2121 on the floor of the inner chamber
(either instead of a pulsation damper 6767 or in front of both
pulsation dampers, see Fig. 14 of 732/733 Manual).
2
Connect suppressor block
733.0010 IC Separation Center (one-channel system):
Remove the plastic stopper from opening 3131 on the rear
panel of IC Separation Center 733 and push cable 2323
mounted on the suppressor block 2121 through this opening.
733.0020 IC Separation Center (two-channel system):
Remove the plastic stopper from opening 4141 or 4343 on the
733 IC Separation Center and push cable 2323 mounted on the
suppressor block 2121 through the opening.
Connect cable 2323 to connection 1717 “Suppressor“ of the 753
Suppressor Module (see Fig. 8).
3
Connect column to injector
see section 2.8.6 of 732/733 Manual
4
Rinse column
see section 2.8.6 of 732/733 Manual
5
Connect column to suppressor
Screw inlet capillary 2424 marked with "Eluent" at suppressor
connection 2222 (see Fig. 6) to outlet end of separating column
7272 using a 6.2744.010 compression fitting.
6
Fix column
Insert one or two column holders 8080 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rails 7979 and fasten separating
column in the column holder.
7
Connect suppressor to detector block
Screw outlet capillary 2929 marked with "Detector" at suppres-
sor connection 2222 to coupling 9292 (6.2620.060) using a
6.2744.010 compression fitting.
Screw inlet capillary 7878 of detector block 7777 to the other end
of the coupling 9292.