User Manual
2 Installation
753 Suppressor Module / 8.753.1003 Instructions for Use
16
1
Insert suppressor block
• 733.0010 IC Separation Center (1-channel system):
Place suppressor block 2121 to the right-hand side of detector
block 7777 (see Fig. 14 of 732/733 Manual) on the floor of the
inner chamber in such a way that suppressor connection 2222
points in the direction of the door.
• 733.0020 IC Separation Center (two-channel system):
Place suppressor block 2121 on the floor of the inner chamber
(either instead of a pulsation damper 6767 or in front of both
pulsation dampers, see Fig. 14 of 732/733 Manual).
2
Connect suppressor block
• 733.0010 IC Separation Center (one-channel system):
Remove the plastic stopper from opening 3131 on the rear
panel of IC Separation Center 733 and push cable 2323
mounted on the suppressor block 2121 through this opening.
• 733.0020 IC Separation Center (two-channel system):
Remove the plastic stopper from opening 4141 or 4343 on the
733 IC Separation Center and push cable 2323 mounted on the
suppressor block 2121 through the opening.
• Connect cable 2323 to connection 1717 “Suppressor“ of the 753
Suppressor Module (see Fig. 8).
3
Connect column to injector
• see section 2.8.6 of 732/733 Manual
4
Rinse column
• see section 2.8.6 of 732/733 Manual
5
Connect column to suppressor
• Screw inlet capillary 2424 marked with "Eluent" at suppressor
connection 2222 (see Fig. 6) to outlet end of separating column
7272 using a 6.2744.010 compression fitting.
6
Fix column
• Insert one or two column holders 8080 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rails 7979 and fasten separating
column in the column holder.
7
Connect suppressor to detector block
• Screw outlet capillary 2929 marked with "Detector" at suppres-
sor connection 2222 to coupling 9292 (6.2620.060) using a
6.2744.010 compression fitting.
• Screw inlet capillary 7878 of detector block 7777 to the other end
of the coupling 9292.