753 Suppressor Module / 8.753.1003 Instructions for Use
14
• Pull outlet capillary 2727 marked with "Waste" at suppressor
connection 2222 (see Fig. 6) through one of the rear panel
openings 2929 of the 690 Ion Chromatograph.
• Lead outlet capillary 2727 to a sufficiently large waste container
and fix in place.
10
Startup of 753 Suppressor Module
• Set switch <REMOTE> 22 on 753 Suppressor Module to
"OFF".
• Switch on 753 Suppressor Module with mains switch 1313 and
set switch <PUMP> 33 to "ON".
• Adjust the contact pressure for both tubing cartridges: press
contact pressure lever 55 upwards until the solutions just start
to be drawn in. Then press the contact pressure lever up-
wards until it clicks once more to obtain optimal contact pres-
sure.
• Before switching the suppressor to the next position for the
first time rinse all three suppressor units with solution for
5 min.
• Check all tubing from the storage vessels through the tubing
cartridges and the suppressor up to the waste containers for
leakage of liquids. If liquid escapes anywhere the corre-
sponding compression fitting must be tightened further or
changed.
Pump tubings are consumable material with a lifetime which depends
on the contact pressure. This is why the tubing cartridges should be
raised completely by loosening snap-action lever 1010 on the right-hand
side if the pump is to remain switched off for a considerable length of
time (the set contact pressure remains unchanged).