2.3 Connection to the 690 Ion Chromatograph
753 Suppressor Module / 8.753.1003 Instructions for Use
13
3030 3232 3333 3434 44 55 3434 3232 25/2825/281010 31313131
Fig. 7: Installing pump tubings
44 Tubing cartridge 3030 PTFE tubing (6.1803.020)
55 Contact pressure lever 3131 PEEK compression fitting
(6.2744.010)
1010 Snap-action lever 3232 Coupling (6.2744.030)
2525 Suppressor inlet capillary
for H
2
SO
4
3333 Pump tubing (6.1826.050)
2828 Suppressor inlet capillary
for H
2
O
3434 Stopper (white-yellow)
9
Suppressor connection 3: H
2
O
Pull the inlet capillary 2828 marked with “H
2
O“ at suppressor
connection 2222 (see Fig. 6) through one of the rear panel
openings 2929 of the 690 Ion Chromatograph or through the
door gap in the right-hand lower corner of the interior door
1313.
Mount a PEEK compression fitting 3131 (6.2744.010) on the
end of inlet capillary 2828 and screw this on to coupling 3232
(6.2744.030) (see Fig. 7).
Push coupling 3232 on to the outlet end of the second pump
tubing 3333 (see Fig. 7).
Mount suction tubing: cut a piece of PTFE Tubing 3030
(6.1803.020) to the desired length.
Mount a PEEK compression fitting 3131 (6.2744.010) on one
end of PTFE tubing 3030 and screw this on to coupling 3232 (see
Fig. 7).
Push coupling 3232 on to the inlet end of the second pump
tubing 3333 (see Fig. 7).
Immerse the other end of the suction tubing in a vessel
containing rinsing solution (normally dist. H
2
O) and fix in
place.