2.3 Connection to the 690 Ion Chromatograph
753 Suppressor Module / 8.753.1003 Instructions for Use
13
3030 3232 3333 3434 44 55 3434 3232 25/2825/281010 31313131
Fig. 7: Installing pump tubings
44 Tubing cartridge 3030 PTFE tubing (6.1803.020)
55 Contact pressure lever 3131 PEEK compression fitting
(6.2744.010)
1010 Snap-action lever 3232 Coupling (6.2744.030)
2525 Suppressor inlet capillary
for H
2
SO
4
3333 Pump tubing (6.1826.050)
2828 Suppressor inlet capillary
for H
2
O
3434 Stopper (white-yellow)
9
Suppressor connection 3: H
2
O
• Pull the inlet capillary 2828 marked with “H
2
O“ at suppressor
connection 2222 (see Fig. 6) through one of the rear panel
openings 2929 of the 690 Ion Chromatograph or through the
door gap in the right-hand lower corner of the interior door
1313.
• Mount a PEEK compression fitting 3131 (6.2744.010) on the
end of inlet capillary 2828 and screw this on to coupling 3232
(6.2744.030) (see Fig. 7).
• Push coupling 3232 on to the outlet end of the second pump
tubing 3333 (see Fig. 7).
• Mount suction tubing: cut a piece of PTFE Tubing 3030
(6.1803.020) to the desired length.
• Mount a PEEK compression fitting 3131 (6.2744.010) on one
end of PTFE tubing 3030 and screw this on to coupling 3232 (see
Fig. 7).
• Push coupling 3232 on to the inlet end of the second pump
tubing 3333 (see Fig. 7).
• Immerse the other end of the suction tubing in a vessel
containing rinsing solution (normally dist. H
2
O) and fix in
place.