User Manual

4.2. Diagnosis
751 GPD Titrino
201
9 Memory card test
Press <9>.
memory card test
Insert memory card.
The memory card has to be formated from a 751 GPD Titrino otherwise the test does not recognize the card
and on the display appears „memory card 000 KB end“.
Press <Enter>.
The test runs automatically. If no fault arises then there appears:
memory card 128 KB o.k.
Press <Clear>.
diagnose press key 0...9
10 Internal dosing unit D 0
Insert exchange unit and put the burette tip into a collecting receptacle.
Press <Clear>.
The Titrino gets out of the diagnosis menu and returns to the user program.
The spindle drive and the cock changeover drive of the internal dispensing unit can be
checked with the aid of the function <prep> (see page 99).
Set the parameters of this function accord. to the following list (for setting the parameters
see page 9).
configuration
>prep.dosing elements
power ON prep: OFF
report: OFF
dos.element: internal D0
warn.interv.D0 OFF min
volume D0 3.5 ml
cycles D0 2
dos.rate DX max. ml/min
fill rate DX max. ml/min
Press <Prep>.
intern DO prep