4.2. Diagnosis
751 GPD Titrino
194
4.2.3 Equipment required:Equipment required:
voltage calibrator, e.g. 1.642.0010 Metrohm pH Simulator
or 1.767.0010 Calibrated Reference for mV, pH, ê æS, °C
highly insulated interconnection cable 6.2108.060
resistor switch-box, class 0.1 % (or resistor 14.3 k, 0.1 %)
cable 3.496.5070
exchange units, if possible with different cylinder volumes (or 3.496.0070 dummy exchange unit)
stop watch or watch with second hand
digital or analogue voltmeter (if need be, connect a calibrated recorder)
2 connecting cables with 4 mm banana plugs
test plug 3.496.8550 (necessary only if plug ‘Remote’ should be checked)
test plug 3.496.8560 (necessary only if plug ‘RS 232’ should be checked)
1 Prepare instruments for diagnostic test
Power off.
Disconnect all external connections (cables at rear, except mains cable and keyboard).
Remove exchange unit.
Power on and immediately press and hold the <9> key until the powerup test pattern dis-
appears.
diagnose press key 0...9
2 Perform display test
Press <2>.
display test
Press <Enter>.
Characters for a visual check of the display are generated on the eight lines.
Test sequence:
a) The display is cleared and overwritten from the top left to the bottom right with a dot
pattern ( ²²²²²²²² ).
b) The display is cleared and overwritten from the top left to the bottom right with a dot
pattern( °°°°°°°°.
c) The display is continuously cleared and overwritten from the top left to the bottom
right with the complete character set (see page 195 / Fig. 1). At the same time with
moving display the LED’s „COND.“, „STATISTICS“ and „SILO“ are swtiches on and
off.
The test sequence can be held and then continued at any time by pressing <5>.
Block 2 is quit by pressing <Clear>.
diagnose press key 0...9