4.1 Error and special messages
751 GPD Titrino
191
system error 14 No communication between the Titrino and the connected
Remote Box.
Possible causes:
. The Remote Box was connected when the Titrino was
running
. Titrino has a fault.
. Remote Box has a fault.
Remedy: Set under <CONFIG>, >peripheral units, "Remote
Box: OFF", switch off Titrino, take away Remote Box and
switch on Titrino. Contact Metrohm service.
temp out of limit The determination has been stopped as a limit value of the
temperature monitoring has been violated.
temp Message in the measuring point list: A limit value of the
temperature monitoring has been violated.
time-out PC keyboard A connected PC keyboard has been used to call up an ad-
dress (e.g. <F12>) and the connection has then been inter-
rupted.
Possible causes:
. Remote Box has a fault.
. PC keyboard has a fault.
Exit: Correct fault and switch Titrino off/on.
TIP terminated TIP has been terminated.
transmission error With a Remote Box connected characters are received which
cannot be interpreted.
Possible causes:
. Wrong key combination has been pressed.
. Wrong PC keyboard has been selected.
. The barcode reader supplies garbled characters.
. The Remote Box has a fault.
Exit: Rectify fault and switch Titrino off/on.
validate instrument Validation interval has elapsed.
Exit: <CLEAR> or new start.
wrong card The card was inserted/removed during an operation with
<CARD>.
Exit: <CLEAR>
wrong sample With SET, KFT or DOC with preset titration direction the first
measured value is outside the end point.