3.2 Remote control commands
751 GPD Titrino
173
3.2.2.3.2.2.150.150. IInfo.As.Assembly.B.Bur.S.Select internal D0internal D0, external D1, external D2
Selection of buret for assembly functions.
3.2.2.3.2.2.151.151. IInfo.As.Assembly.B.Bur.E.Empty $G, $S, $H, $C
Starts the function "empty". Only possible with the burette type "Dosino". The
parameters are under the function &Config.DosPrep, see 3.2.2.118 up to
3.2.2.120.
3.2.2.3.2.2.152.152. IInfo.As.Assembly.B.Bur.P.Prep $G, $S, $H, $C
Starts the function "preparation". The parameters are under the function
&Config.DosPrep, see 3.2.2.110 up to 3.2.2.120.
3.2.2.3.2.2.153.153. AAssembly.B.Bur.R.Rates.F.Forward.S.Selected digital, analog analog
AAssembly.B.Bur.R.Rates.F.Forward.D.Digital 0...150, max.max.
AAssembly.B.Bur.R.Rates.R.Reverse.S.Selected digital, analog analog
AAssembly.B.Bur.R.Rates.R.Reverse.D.Digital 0...150, max.max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in cur-
rent use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.3.2.2.154.154. AAssembly.B.Bur.F.Fill $G, $H, $C
$G starts the 'FILL' mode of the burette function.
3.2.2.3.2.2.155.155. AAssembly.B.Bur.M.ModeDis $G, $S, $H, $C
AAssembly.B.Bur.M.ModeDis.S.Selected volumevolume, time
AAssembly.B.Bur.M.ModeDis.V.V 0.0001...0.10.1...9999
AAssembly.B.Bur.M.ModeDis.T.Time 0.25...11...86400
AAssembly.B.Bur.M.ModeDis.VS.VStop 0.0001...9999, OFF OFF
AAssembly.B.Bur.M.ModeDis.A.AutoFill ON, OFF OFF
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop: Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
3.2.2.3.2.2.156.156. AAssembly.M.Meas.S.Status ON, OFF OFF
AAssembly.M.Meas.M.MeasInput 11, 2, Diff., Ipol, Upol, Temp
AAssembly.M.Meas.I.Ipol ±127...11...+127
AAssembly.M.Meas.U.Upol ±1270...400400...+1270