3.2. Remote control commands
751 GPD Titrino
172
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with DET, MET, SET, KFT.
.MeasPt.X"165" Volume (DET, MET), time (STAT, DOS, DOC, SET, KFT,
MEAS), resp. of the MPList in s
.MeasPt.Y"3.654" Measured value (DET, MET), volume (STAT, DOS, DOC, SET,
KFT), resp. of the MPList in mL
.MeasPt.Z1"6.34" Measured value (STAT, DOS, DOC, SET, KFT, MEAS) of the
MPList, format depends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.3.2.2.146.146. IInfo.A.ActualInfo.Ov.Oven.H.HeatTime read only
IInfo.A.ActualInfo.Ov.Oven.S.SampleTemp read only
IInfo.A.ActualInfo.Ov.Oven.L.LowTemp read only
IInfo.A.ActualInfo.Ov.Oven.H.HighTemp read only
IInfo.A.ActualInfo.Ov.Oven.G.GasFlow read only
IInfo.A.ActualInfo.Ov.Oven.U.UnitFlow read only
$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.
3.2.2.3.2.2.147.147. IInfo.A.ActualInfo.D.Display.L.L1 up to 32 ASCII characters
IInfo.A.ActualInfo.D.Display.L8.L8 up to 32 ASCII characters
IInfo.A.ActualInfo.D.Display.D.DelAll $G
Lines of the display. The display can be written to from the computer. Proceed
as follows:
1. Lock the display, see 3.2.2.163.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.163
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.105) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.3.2.2.148.148. IInfo.A.ActualInfo.C.Comport.N.Number read only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.3.2.2.149.149. IInfo.As.Assembly.C.CycleTime read only
IInfo.As.Assembly.E.ExV read only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.