User Manual
3.2. Remote control commands
751 GPD Titrino
170
C24.Name: Name of the assigned value "RS1"
C24.Value: Value "2.222"
C24.Unit: Unit of the assigned value "%"
C26.ActN: Number of single results "3"
C26.Mean: Mean (decimal places as for the result itself) "3.421"
C26.Std: Standard deviation (decimal places as for the result + 1) "0.0231"
C26.RelStd: Relative standard deviation (in %, 2 decimal places) "0.14"
3.2.2.3.2.2.140.140. IInfo.A.ActualInfo.I.Inputs.S.Status read only
IInfo.A.ActualInfo.I.Inputs.C.Change read only
IInfo.A.ActualInfo.I.Inputs.Cl.Clear $G
IInfo.A.ActualInfo.O.Outputs.S.Status read only
IInfo.A.ActualInfo.O.Outputs.C.Change read only
IInfo.A.ActualInfo.O.Outputs.Cl.Clear $G
Status sends the current status of the I/O lines, Change sends the information
regarding whether a change in status of a line has taken place since the last
clearing, Clear clears the change information. For the output, there is a conver-
sion from binary to decimal, e.g.
| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
Line No. | 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Output: 2
1
+ 2
3
= "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 221ff):
InputsInputs: OutputsOutputs:
0 Start (pin 21) 0 Ready (pin 5)
1 Stop (pin 9) 1 Cond. ok (pin 18)
2 Enter (pin 22) 2 Titration (pin 4)
3 Clear (pin 10) 3 EOD (pin 17)
4 Smpl Ready (pin 23) 4 Monitoring, line L4 (pin 3)
5 pin 11 5 Error (pin 16)
6 pin 24 6 Activate, line L6 (pin 1)
7 pin 12 7 Pulse for recorder (pin 2)
8 not used (pin 6)
9 not used (pin 7)
10 Monitoring (pin 8)
11 Monitoring (pin 13)
12 Monitoring (pin 19)
13 Monitoring (pin 20)
3.2.2.3.2.2.141.141. IInfo.A.ActualInfo.A.Assembly.C.CyclNo read only
$Q sends the current cycle number of the voltage measurement cycle, e.g.
"127". From the cycle number and the cycle time (see 3.2.2.149), a time frame
can be set up.
The cycle number is set to 0 on switching on the instrument, on every start and
for QuickMeas. It is incremented as long as the instrument remains switched
on.
3.2.2.3.2.2.142.142. IInfo.A.ActualInfo.A.Assembly.Co.Counter.V.V read only
IInfo.A.ActualInfo.A.Assembly.Co.Counter.C.Clear $G