User Manual
3.2. Remote control commands
751 GPD Titrino
168
IInfo.D.DetermData.E.ExV read only/read+write
IInfo.D.DetermData.M.MPList.1.X.1.X read only/read+write
IInfo.D.DetermData.M.MPList.1.Y.1.Y read only/read+write
IInfo.D.DetermData.M.MPList.1.Z.1.Z1 read only/read+write
IInfo.D.DetermData.M.MPList.1.Z2.1.Z2 read only/read+write
for every measuring point
Determination data in hexadecimal format. A measuring point list is available in
mode DET, MET, STAT, DOS, DOC, SET, KFT, and MEAS.
Recalculation of the measuring data is triggered with $G.
.Write: With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.
.ExV: Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
.MPList.Z1: Z1 coordinate, measuring value
.MPList.Z2: Z2 coordinate, temperature
3.2.2.3.2.2.135.135. IInfo.T.TitrResults.R.RS.1.V.1.Value read only
etc., up to .9.9
IInfo.T.TitrResults.E.EP.1.V.1.V read only
IInfo.T.TitrResults.E.EP.1.M.1.Meas read only
etc., up to .2.2
IInfo.T.TitrResults.V.Var.C.C40 read only/read+write
etc., up to .C47.C47
IInfo.T.TitrResults.V.Var.DTime read only/read+write
.RS.RS: Values of the calculated results.
.EP.EP: Endpoints with DET, MET, SET, KFT:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in pH "5.12", mV (with U and Ipol) "-
241" or uA (with Upol) "43.7".
..VarVar: Various variables. You may overwrite the variables C40...C45, see
3.2.2.134.
C40: Initial measured value in pH "5.12", mV (with U and Ipol) "41",
uA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final measured
value.
C41: End volume with SET, STAT, DOS, DOC in ml, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of a SET/KFT titration from the condi-
tioning in ul/min, "3.5".
C44: (Last measured) temperature in °C. Used for the temperature
compensation in pH measurements.
C45: Start volume with SET, STAT, DOC in ml, "2.800".
C46: Asymmetry pH of CAL, "6.89".
C47: Relative electrode slope of CAL, "0.9950".
C48: Voltage at maximum voltage (not valid in CAL and TIP)
C49: Voltage at minimum voltage (not valid in CAL and TIP)
DTime: Dosing time in s in DOS/DOC or time for the drift correction
in SET/KFT with conditioning, resp.