3.2 Remote control commands
751 GPD Titrino
163
3.2.2.3.2.2.107.107. CConfig.A.Aux.P.Prog read only
Output of the program version.
The Titrino sends "751.0020" on requests with $Q.
3.2.2.3.2.2.108.108. CConfig.R.RSSet1 $G
CConfig.R.RSSet1.B.Baud 300, 600, 1200, 2400, 4800, 9600, 9600,
19200, 38400, 57600, 115200
CConfig.R.RSSet1.D.DataBit 7, 88
CConfig.R.RSSet1.S.StopBit 11, 2
CConfig.R.RSSet1.P.Parity even, odd, none none
CConfig.R.RSSet1.H.Handsh HWsHWs, SWchar, SWline, none
$G sets all RS settings. The changes are performed only if the instrument is in-
active. After the setting of the interface parameters, wait at least 2 s to allow the
components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud rate,
data bit, stop bit, parity and type of handshake, see also page 180 ff. Baud
rates >9600 need a PC which is equipped accordingly (e.g. with 16550 com-
ponent).
The setting of the values must be initiated with $G immediately after entry of the
values.
3.2.2.3.2.2.109.109. CConfig.C.ComVar.C30C30
with up to .C39.C39, etc. 0... +999 999
Values of the common variables from C30 up to C39. Insert the common vari-
ables directly or describe the determination results directly from the method,
see 3.2.2.7
3.2.2.3.2.2.110.110. CConfig.D.DosPrep.PPowerOnPrep ON, OFFOFF
Warning after power on, that the preparation of the buret has to be executed.
3.2.2.3.2.2.111.111. CConfig.D.DosPrep.RReport ON, OFFOFF
Report output after preparation of burets.
3.2.2.3.2.2.112.112. CConfig.D.DosPrep.SSelect internal D0internal D0, external D1, external D2
Selection of buret.
3.2.2.3.2.2.113.113. CConfig.D.DosPrep.DX.WDX.WarnInterval 5...9999, OFFOFF
Time interval in min for automatic warning that a preparation has to be executed.
3.2.2.3.2.2.114.114. CConfig.DDosPrep.DXDX(.D.Dosimat).VV 0...3.53.5...99999.99
Volume in ml, which will be expelled at the preparation. Start of sequence see
3.2.2.152.