User Manual
3.2 Remote control commands
751 GPD Titrino
161
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.3.2.2.92.92. CConfig.M.Monitoring.C.Calibration.S.Status ON, OFFOFF
CConfig.M.Monitoring.C.Calibration.M.MeasInput 11, 2, diff
CConfig.M.Monitoring.C.Calibration.I.Interval 1...77...9999
CConfig.M.Monitoring.C.Calibration.C.Counter 00...9999
Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval: Time interval in days for calibration from the last calibration date.
.Counter: Time counter in days since last calibration. The counter is reset to
zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.
3.2.2.3.2.2.93.93. CConfig.M.Monitoring.S.Service.S.Status ON, OFFOFF
CConfig.M.Monitoring.S.Service.D.Date XXXX-XX-XX
Monitoring of service interval.
3.2.2.3.2.2.94.94. CConfig.M.Monitoring.D.DiagRep ON, OFFOFF
Printing of system test report after each switching on of the Titrino.
3.2.2.3.2.2.95.95. CConfig.P.PeriphUnit.C.CharSet1 Epson, Seiko, Citizen, HP, IBM IBM
CConfig.P.PeriphUnit.CharSet2.CharSet2
Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.3.2.2.96.96. CConfig.P.PeriphUnit.R.RepToComport 11, 2, 1&2
Selection of COM of the Titrino where manually triggered reports should be out-
putted.
3.2.2.3.2.2.97.97. CConfig.P.PeriphUnit.B.Balance SartoriusSartorius,Mettler,Mettler AT,
AND,Precisa
Selection of the balance type.
3.2.2.3.2.2.98.98. CConfig.P.PeriphUnit.S.Stirrer ON, OFFOFF
Automatic stirrer control. With "ON" the stirrer will be switched on after starting
a method. At the end of the method it is switched off again. Valid for DET, MET,
SET, KFT, STAT, DOC.
3.2.2.3.2.2.99.99. CConfig.P.PeriphUnit.Rem.RemoteBox.S.Status ON, OFFOFF