3.2.2.3.2.2.70.70. MMode.P.Parameter.C.Calibration.S.SignalDrift 0.5...22...999, OFF
MMode.P.Parameter.C.Calibration.E.EquTime 0...110110...9999, OFF
Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding crite-
rion is switched off. If both criterions are on OFF, the measured value is ac-
quired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 20. After it has been edited once, it
remains in force with the set value.
3.2.2.3.2.2.71.71. MMode.P.Parameter.C.Calibration.El.ElectrodeId up to 8 ASCII char
Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.131.
3.2.2.3.2.2.72.72. MMode.P.Parameter.C.Calibration.Sm.SmplChanger ON, OFF OFF
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.3.2.2.73.73. MMode.P.Parameter.C.Calibration.A.ActPulse first, all, OFF OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page 224.
3.2.2.3.2.2.74.74. MMode.P.Parameter.S.Sequence.X.S.X.Select method, pause, L4 output,
L6 output, info, prep, stirrer, OFF OFF
Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.75.
3.2.2.3.2.2.75.75. MMode.P.Parameter.S.Sequence.X.M.X.Method Method name
MMode.P.Parameter.S.Sequence.X.P.X.Pause 00...999 999, INF
MMode.P.Parameter.S.Sequence.X.L.X.L4Output active, inactive,
pulse, OFF OFF
MMode.P.Parameter.S.Sequence.X.L6.X.L6Output as for L4
MMode.P.Parameter.S.Sequence.X.I.X.Info up to 16 ASCII characters
MMode.P.Parameter.S.Sequence.X.Pr.X.Prep internal D0internal D0, external D1,
external D2
MMode.P.Parameter.S.Sequence.X.S.X.Stirrer ONON, OFF
Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the
card. Up to 8 ASCII characters.
.Pause: Pause time in s. INF means infinite. Continue the sequence with &m
$G.
.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin
3) in a submethod sets an output set to active in TIP to inactive.