3.2 Remote control commands
751 GPD Titrino
153
wait: Interrupt dosing until the limit value is no longer violated, then con-
tinue automatically.
3.2.2.3.2.2.50.50. MMode.P.Parameter.M.Monitoring.L.L10Output.A.AssignOutput
meas, temp, rate, all, none none
MMode.P.Parameter.M.Monitoring.L.L10Output.L.Limit upper,lower,anyany
MMode.P.Parameter.M.Monitoring.L.L10Output.O.Output active, pulsepulse
Parameters for STAT, DOS, DOC: Output of a signal on L10 output (pin 8) of
the remote socket on limit value violation.
The output can be assigned a definite or all monitoring value(s) as well as the
violated limit and the type of signal to be outputted determined.
3.2.2.3.2.2.51.51. MMode.P.Parameter.P.Presel.R.RateDisp ON, OFFOFF
Parameters for STAT, DOC: Display of the current rate during the determination.
3.2.2.3.2.2.52.52. MMode.P.Parameter.D.DosPara.T.Type volumevolume, time, rate
MMode.P.Parameter.D.DosPara.V.Volume.V.Volume 0.000...1010...99999.99
MMode.P.Parameter.D.DosPara.V.Volume.D.DisType time, rate rate
MMode.P.Parameter.D.DosPara.V.Volume.R.Rate 0.001...150, max.max.
MMode.P.Parameter.D.DosPara.V.Volume.T.Time 1...100100...999 999
MMode.P.Parameter.D.DosPara.Ti.Time.T.Time 1...100100...999 999
MMode.P.Parameter.D.DosPara.Ti.Time.D.DisType volumevolume, rate
MMode.P.Parameter.D.DosPara.Ti.Time.R.Rate 0.001...150, max.max.
MMode.P.Parameter.D.DosPara.Ti.Time.V.Volume 0.000...1010...99999.99
MMode.P.Parameter.D.DosPara.R.Rate.R.Rate 0.001...150, max.max.
MMode.P.Parameter.D.DosPara.R.Rate.S.StopType volumevolume, time
MMode.P.Parameter.D.DosPara.R.Rate.T.Time 1...100100...999 999
MMode.P.Parameter.D.DosPara.R.Rate.V.Volume 0.000...1010...99999.99
Parameters for DOS: In the equation Rate = Volume/Time, two parameters are
preset, the third is calculated. Depending on the entry of the type, the following
information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.
3.2.2.3.2.2.53.53. MMode.P.Parameter.D.DosPara.P.Pause 00...999 999
Parameters for DOS: Pause time in s.
3.2.2.3.2.2.54.54. MMode.P.Parameter.D.DosPara.TD.TDelta 1...1010...999 999
Parameters for DOS, DOC: Time interval for the entry of a measured point in the
list of measured points.
3.2.2.3.2.2.55.55. MMode.P.Parameter.D.DosPara.Te.Temp -170.0...25.025.0...500.0
Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.