User Manual
3.2 Remote control commands
751 GPD Titrino
151
If an Oven is connected, its results will be incorporated into the result report of
the Titrino.
If there is no Oven connected via RS232, this parameter has to be on "no".
3.2.2.3.2.2.40.40. MMode.P.Parameter.C.CtrlPara.Mi.MinRate 0.01...25.025.0...9999
Parameter for STAT: Control parameters.
.MinRate: Minimum titration rate in ul/min.
3.2.2.3.2.2.41.41. MMode.P.Parameter.T.TitrPara.T.TStart 00...999 999
MMode.P.Parameter.T.TitrPara.StartM.StartMeas pH: 0.00...±20.00, OFF OFF
U, Ipol: 0...±2000, OFF OFF
Upol: 0.0...±200.0, OFF OFF
MMode.P.Parameter.T.TitrPara.R.RateStart 0.01...150, OFFOFF
Parameters for STAT: Start of the measured point acquisition after the start
condition has been met.
.TStart: Start time in s.
.StartMeas: Start measured value.
.RateStart: Start rate in mL/min. The current rate must be less than the start
rate .This criterion does not become effective until 10 s after the
start.
3.2.2.3.2.2.42.42. MMode.P.Parameter.St.StopCond.T.TStop.T.Type abs.,rel.,delta,delay,OFF,OFF
MMode.P.Parameter.St.StopCond.T.TStop.T.Time 0...999 999999 999
MMode.P.Parameter.St.StopCond.T.TStop.F.Factor 0...±999 99999 99
MMode.P.Parameter.St.StopCond.T.TStop.D.Delta 0...999 999999 999
MMode.P.Parameter.St.StopCond.T.TStop.Dela.Delay 0...999 999999 999
Parameters for STAT: Stop time.
If an absolute stop time (abs.) has been selected, the time is in s (Time).
A relative stop time (rel.) is calculated as a function of the sample size:
Stop time in s = sample size * factor
The factor selected (Factor) applies.
delta: Time in s after control point is reached for the first time.
delay: Time in s after last dosing step.
OFF means that the criterion is not monitored.
3.2.2.3.2.2.43.43. MMode.P.Parameter.St.StopCond.R.RateStop 0.01...150, OFF OFF
Parameters for STAT: Stop rate in ml/min. The current rate must be less than
the start rate. This criterion does not become effective until 10 s after the start.
3.2.2.3.2.2.44.44. MMode.P.Parameter.E.Evaluation.T.TimeWin.1.L.1.LowLim 0...999 999,OFFOFF
MMode.P.Parameter.E.Evaluation.T.TimeWin.1.U.1.UpLim 0...999 999, OFFOFF
Parameters for STAT: Evaluation of rates in time windows. Entry of the times in
s. The evaluated rates are available as variables C8X (X = 1...9).
3.2.2.3.2.2.45.45. MMode.P.Parameter.E.Evaluation.F.FixVol.1.V.1.Value 0...999 999, OFFOFF