User Manual

3.2. Remote control commands
751 GPD Titrino
150
Parameter for SET, KFT: Pause time in s. Runs before dosing the start volume.
3.2.2.3.2.2.34.34. MMode.P.Parameter.T.TitrPara.E.ExtrT 00...999 999
Parameter for SET, KFT: Extraction time in s.
3.2.2.3.2.2.35.35. MMode.P.Parameter.T.TitrPara.TD.TDelta 1...22...999 999
Parameter for SET, KFT, STAT: Time interval in s for the entry of a measure-
ment point in the list of measured points.
3.2.2.3.2.2.36.36. MMode.P.Parameter.P.Presel.C.Cond ON, OFF OFF
MMode.P.Parameter.P.Presel.DDriftDisp ONON, OFF
MMode.P.Parameter.P.Presel.DCDCor.T.Type auto, man., OFF OFF
MMode.P.Parameter.P.Presel.DCDCor.V.Value 0.00.0...99.9
Parameters for SET, KFT:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2.2.3.2.2.37.37. MMode.P.Parameter.C.CtrlPara.E.EP pH: 0...±20.00, OFF OFF
U, Ipol: 0...±2000, OFF OFF
Upol: 0...±200.0, OFF OFF
MMode.P.Parameter.C.CtrlPara.U.UnitEp read only
Parameters for KFT, STAT: Setting of the EP, resp. Control point as a pH value,
in mV (with U and Ipol) or uA (with Upol). The corresponding unit can be read
with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 uA.
3.2.2.3.2.2.38.38. MMode.P.Parameter.C.CtrlPara.D.Dyn pH: 0.01...20.00, OFF OFF
U, Ipol: 1...2000, OFF OFF
Upol: 0.1...200.0, OFF OFF
MMode.P.Parameter.C.CtrlPara.UnitD.UnitDyn read only
MMode.P.Parameter.C.CtrlPara.M.MaxRate 0.01...1010...150, max.
MMode.P.Parameter.C.CtrlPara.Mi.MinIncr 0.1...9.9, min. min.
Parameters for KFT, STAT: Control parameters.
.Dyn: Dynamics (control range) in pH, mV (with U and Ipol) or uA (with Upol).
The corresponding unit can be read with .UnitDyn. Standard with
KFT Ipol 100, with KFT Upol 10.
.MaxRate: Maximum allowed titration rate in ml/min. Max. means maximum
possible rate with mounted Exchange Unit.
.MinIncr: Parameter for KFT. Minimum volume increment in ul.
3.2.2.3.2.2.39.39. MMode.P.Parameter.P.Presel.O.Oven COM1, COM2, nono
Parameter for KFT.