User Manual

3.2. Remote control commands
751 GPD Titrino
148
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.3.2.2.25.25. MMode.P.Parameter.E.Evaluation.EEPC DET: 0...55...200
MET pH: 0.1...0.500.50...9.99
U, Ipol: 1...3030...999
Upol: 0.1...22...99.9
MMode.P.Parameter.E.Evaluation.R.Recognition.S.Selected allall, greatest,
last, window, OFF
MMode.P.Parameter.E.Evaluation.R.Recognition.W.Window.1.L.1.LowLim
pH: 0...±20.00, OFF OFF
U, Ipol: 0...±2000, OFF OFF
Upol: 0...±200.0, OFF OFF
MMode.P.Parameter.E.Evaluation.R.Recognition.W.Window.1.U.1.UpLim
Input range as LowLim
etc. up to 9 windows
MMode.P.Parameter.E.Evaluation.F.FixEP.1.V.1.Value pH: 0...±20.00, OFF OFF
U; Ipol: 0...±2000, OFF OFF
etc. up to 9 fix EP's Upol: 0...±200.0, OFF OFF
MMode.P.Parameter.E.Evaluation.p.pK ON, OFFOFF
Parameters for DET and MET: Evaluation of the EP's, see page 21.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U and
Ipol) or uA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 23.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.3.2.2.26.26. MMode.P.Parameter.P.Presel.I.IReq id1, id1&2, all, OFF OFF
MMode.P.Parameter.P.Presel.S.SReq value, unit, all, OFF OFF
Parameters for DET, MET, SET, KFT, STAT, DOS, DOC, MEAS: Automatic in-
quiry after the start of the determination. From such an inquiry, the determina-
tion continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.122) or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.3.2.2.27.27. MMode.P.Parameter.P.Presel.L.LimSmplSize.S.Status ON, OFFOFF
MMode.P.Parameter.P.Presel.L.LimSmplSize.L.LoLim 0.00.0...999 999
MMode.P.Parameter.P.Presel.L.LimSmplSize.U.UpLim 0.0...999 999999 999
Limit control for the sample size.