User Manual
3.2 Remote control commands
751 GPD Titrino
139
&Assembly
Object Description Input range Reference
& Root
:
à Assembly Assembly control
à . Bur Buret
³ Ã .Select Selection of buret internal D0, external D1
³ ³ external D2 3.2.2.150.
³ Ã .Empty Empties the buret $G,$S,$H,$C 3.2.2.151.
³ Ã .Prep Prepares the buret $G,$S,$H,$C 3.2.2.152.
³ Ã .Rates Rates
³ ³ Ã .Forward Forward rate
³ ³ ³ Ã .Select Type of rate control digital, analog 3.2.2.153.
³ ³ ³ Ã .Digital Digital rate 0...150, max. ditto
³ ³ Ã .Reverse as for forward rate
³ ³ ³ Ã .Select Type of rate control digital, analog ditto
³ ³ ³ Ã .Digital Digital rate 0...150, max. ditto
³ Ã .Fill Fill $G,$H,$C 3.2.2.154.
³ Ã .ModeDis Dispensing $G,$S,$H,$C 3.2.2.155.
³ ³ Ã .Select Type of dispensing control volume, time ditto
³ ³ Ã .V Volume to be dispensed 0.0001...0.1...9999 ditto
³ ³ Ã .Time Time to dispense 0.25...1...86 400 ditto
³ ³ Ã .VStop Limit volume 0.0001...9999, OFF ditto
³ ³ Ã .AutoFill Filling after each increment ON, OFF ditto
³
à .Meas Measuring
³ Ã .Status Measuring ON/OFF ON, OFF 3.2.2.156.
³ Ã .MeasInput Selection of measuring input 1, 2, diff., Ipol, Upol, Temp ditto
³ Ã .Ipol Polarization current 0...1...±127 ditto
³ Ã .Upol Polarization voltage 0...400...±1270 ditto
³
à .Outputs I/O outputs
³ Ã .AutoEOD Automatic output of EOD ON, OFF 3.2.2.157.
³ Ã .SetLines Set I/O lines $G ditto
³ ³ Ã .L0 Signal on L0 active,inactive,pulse,OFF ditto
³ ³ Ã up to L13
³ Ã .ResetLines Reset I/O lines $G ditto
³
à .Stirrer Stirrer ON, OFF 3.2.2.158.
³