User Manual

3.2. Remote control commands
751 GPD Titrino
128
*Parameter Tree part "Parameters for MEAS"
à .Measuring Measuring parameters
³ Ã .SignalDrift Drift for meas.value acquisition depends on meas.quant. 3.2.2.63.
³ Ã .UnitSigDrift Unit of measured value drift read only ditto
³ Ã .EquTime Equilibrium time 0...9999, OFF ditto
³ Ã .MeasInput Measuring input 1, 2, diff. 3.2.2.64.
³ Ã .Ipol Polarization current 0...1...±127 ditto
³ Ã .Upol Polarization voltage 0...400...±1270 ditto
³ Ã .PolElectrTest Test for polarized electrodes ON, OFF ditto
³ Ã .Temp Titration temperature -170.0...25.0...500.0 3.2.2.65.
³ Ã .TDelta Time interv.for meas.acquisition 1...2...999 999 3.2.2.66.
³ ³
à .Statistics Statistics
³ Ã .Status Status of statistics calculation ON, OFF 3.2.2.24.
³ Ã .MeanN No. of individual determinations 2...20 ditto
³ Ã .ResTab Result table
³ ³ Ã .Select original,delete n,delete all ditto
³ ³ Ã .DelN Deletion of individual results 1...20 ditto
³ ³
à .Presel Preselections
³ Ã .IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.26.
³ Ã .SReq Request of sample size after start value, unit, all, OFF ditto
³ Ã .LimSmplSize Limits for sample size 3.2.2.27
³ ³ Ã .Status Status of limit control ON, OFF ditto
³ ³ Ã .LoLim Lower limit 0.0...999 999 ditto
³ ³ Ã .UpLim Upper limit 0.0...999 999 ditto
³ Ã .ActPulse Output of a pulse ON, OFF 3.2.2.28.
*Parameter Tree part "Parameters for CAL"
à .Calibration Calibration parameters
³ Ã .MeasInput Measuring input 1, 2, diff. 3.2.2.67.
³ Ã .CalTemp Calibration temperature -20.0...25.0...120.0 3.2.2.68.
³ Ã .Buffer
³ ³ Ã .1
³ ³ ³ Ã .Value pH value of buffer 1 0...7.00...±20.00 3.2.2.69.
³ ³ Ã .2
³ ³ ³ Ã .Value pH value of buffer 2 0...4.00...±20.00, OFF ditto
³ ³ Ã : up to 9 buffers
³ Ã .SignalDrift Drift for meas.value acquisition depends on meas.quant. 3.2.2.70.
³ Ã .EquTime Equilibrium time 0...26...9999, OFF ditto
³ Ã .ElectrodeId Electrode identification 8 ASCII char. 3.2.2.71.
³ Ã .SmplChanger Calibration on a Titrino ON, OFF 3.2.2.72.
³ Ã .ActPulse Output of a pulse first, all, OFF 3.2.2.73.
³ ³
à .Statistics Statistics
³ Ã .Status Status of statistics calculation ON, OFF 3.2.2.24.
³ Ã .MeanN No.of individual determinations 2...20 ditto
³ Ã .ResTab Result table
³ ³ Ã .Select original,delete n,delete all ditto
³ ³ Ã .DelN Deletion of individual results 1...20 ditto