User Manual

3.2. Remote control commands
751 GPD Titrino
126
*Parameter Tree part "Parameters for DOC"
à .DosPara Dosing parameters
³ Ã .BeginMeas Begin of measuring depends on meas.quant. 3.2.2.57.
³ Ã .EndMeas End of measuring depends on meas.quant. ditto
³ Ã .UnitMeas Unit of measured quantity read only ditto
³ Ã .SweepTime Sweep time 0...300...999 999 3.2.2.58.
³ Ã .Dyn Dynamics depends on meas.quant. 3.2.2.59.
³ Ã .UnitDyn Unit of dynamics read only ditto
³ Ã .MaxRate Maximum dosing rate 0.01... 150, max. ditto
³ Ã .MinRate Minimum dosing rate 0.01...5.0...9999 ditto
³ Ã .Direction Direction +, -, auto 3.2.2.60.
³ Ã .StartV Start volume
³ ³ Ã .Type Type of start volume abs., rel., OFF 3.2.2.61.
³ ³ Ã .V Volume for absolute start volume 0...999.99 ditto
³ ³ Ã .Factor Factor for relative start volume 0...±999 999 ditto
³ ³ Ã .Rate Dispensing rate for start volume 0.01...150.0, max. ditto
³ Ã .Pause Waiting time after start volume 0...999 999 ditto
³ Ã .TDelta Time interv. for meas.acquisition 1...10...999 999 3.2.2.52.
³ Ã .DosUnit Selection of the dosing unit internal D0, external D1,
³ ³ external D2 3.2.2.17.
³ Ã .MeasInput Measuring input 1, 2, diff. 3.2.2.62.
³ Ã .Ipol Polarization current 0...1...±127 ditto
³ Ã .Upol Polarization voltage 0...400...±1270 ditto
³ Ã .PolElectrTest Test for polarized electrodes ON, OFF ditto
³ Ã .Temp Titration temperature -170.0...25.0...500.0 3.2.2.53.
³ ³
à .StopCond Stop conditions
³ Ã .VStopStop volume
³ ³ Ã .Type Type of stop volume abs., rel., OFF 3.2.2.20.
³ ³ Ã .V Volume for absolute stop volume 0...99.99...9999.99 ditto
³ ³ Ã .Factor Factor for relative stop volume 0...±999 999 ditto
³ Ã .FillRate Filling rate 0.01...150.0, max. 3.2.2.23.
³ ³
à .StatisticsStatistics
³ Ã .Status Status of statistics calculation ON, OFF 3.2.2.24.
³ Ã .MeanN No. of individual determinations 2...20 ditto
³ Ã .ResTab Result table
³ ³ Ã .Select original,delete n,delete all ditto
³ ³ Ã .DelN Deletion of individual results 1...20 ditto
³ ³
à .Monitoring Monitoring
³ Ã .MeasVal Monitoring of measured values
³ ³ Ã.Status Status ON, OFF 3.2.2.45.
³ ³ Ã.LowLim Lower limit depends on meas.quant. ditto
³ ³ Ã.UpLim Upper limit depends on meas.quant. ditto
³ ³ Ã.Action Action if out of limits end, hold, wait, none ditto
³ Ã .Temp Monitoring of temperatures
³ ³ Ã.Status Status ON, OFF 3.2.2.47.
³ ³ Ã.LowLim Lower limit -170.0...500.0 ditto
³ ³ Ã.UpLim Upper limit -170.0...500.0 ditto
³ ³ Ã.Action Action if out of limits end, hold, wait, none ditto