User Manual

3.2 Remote control commands
751 GPD Titrino
123
*Parameter Tree part "Parameters for STAT", continuation
à .Evaluation Evaluation
³ Ã .TimeWin Time windows for rate evaluation
³ ³ Ã .1 up to 9 windows
³ ³ ³ Ã.LowLim Lower limit window 1 0...999 999, OFF 3.2.2.44.
³ ³ ³ Ã.UpLim Upper limit window 1 0...999 999, OFF ditto
³ Ã .FixVolFix volumes
³ ³ Ã .1 up to 9 fix volumes
³ ³ ³ Ã.Value Value for fix volume 1 0...999 999, OFF 3.2.2.45.
³ Ã .FixTime Fix times
³ ³ Ã .1 up to 9 fix times
³ ³ ³ Ã.Value Value for fix time 1 0.01...1, OFF 3.2.2.46.
³ ³
à .Monitoring Monitoring
³ Ã .MeasVal Monitoring of measured values
³ ³ Ã.Status Status ON, OFF 3.2.2.47.
³ ³ Ã.LowLim Lower limit depends on meas.quant. ditto
³ ³ Ã.UpLim Upper limit depends on meas.quant. ditto
³ ³ Ã.Action Action if out of limits end, hold, wait, none ditto
³ Ã .Rate Monitoring of rates
³ ³ Ã.Status Status ON, OFF 3.2.2.48.
³ ³ Ã.LowLim Lower limit 0.000...150 ditto
³ ³ Ã.UpLim Upper limit 0.000...150 ditto
³ ³ Ã.Action Action if out of limits end, hold, wait, none ditto
³ Ã .Temp Monitoring of temperatures
³ ³ Ã.Status Status ON, OFF 3.2.2.49.
³ ³ Ã.LowLim Lower limit -170.0...500.0 ditto
³ ³ Ã.UpLim Upper limit -170.0...500.0 ditto
³ ³ Ã.Action Action if out of limits end, hold, wait, none ditto
³ Ã .L10Output Assignment of output L10
³ ³ Ã .AssignOutput To type of limit violation meas,temp,rate,all,none 3.2.2.50.
³ ³ Ã .Limit Assign output to violated limit upper, lower, any ditto
³ ³ Ã .Output Signal to be set on output L10 active, pulse ditto
³ Ã .L11Output as for output L10
³ Ã .L12Output as for output L10
³ Ã .L13Output as for output L10
³ ³
à .Presel Preselections
³ Ã .IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.26.
³ Ã .SReq Request of sample size after start value, unit, all, OFF ditto
³ Ã .LimSmplSize Limits for sample size 3.2.2.27
³ ³ Ã .Status Status of limit control ON, OFF ditto
³ ³ Ã .LoLim Lower limit 0.0...999 999 ditto
³ ³ Ã .UpLim Upper limit 0.0...999 999 ditto
³ Ã .RateDisp Display of drift during cond. ON, OFF 3.2.2.51.
³ Ã .ActPulse Output of a pulse ON, OFF 3.2.2.28.