User Manual
3.1. General rules
751 GPD Titrino
114
E171 Rate too low in DOS. No dispensing possible with the Exchange Unit currently
mounted.
Exit: The error message disappears on next start or &m $S.
E172 In TIP or DOS a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E173 The warning interval of the internal buret D0 called.
Exit: Execute prep &a.b.p $G or start next method.
E174 The warning interval of the external buret D1 called.
Exit: Execute prep &a.b.p $G or start next method.
E175 The warning interval of the external buret D2 called.
Exit: Execute prep &a.b.p $G or start next method.
E176 The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was interrupted
manually.
Exit. The error message disappears on next start.
E177 Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
E178 The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
E180 Memory card write-protected.
Exit: Send new command.
E181 Memory card not formatted.
Exit: Send new command.
E182 Memory card not accessible.
Exit: Send new command.
E183 A directory with the same name exists already on the memory card.
Exit: Send new command.
E184 Measured value below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E185 Measured value above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E186 Temperature below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E187 Temperature above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E188 Rate below lower limit.
Exit: The error message disappears when again within the limits or &m $S.
E189 Rate above upper limit.
Exit: The error message disappears when again within the limits or &m $S.
E196 Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
E197 Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
E198 Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
E199 Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.