3.1 General rules
751 GPD Titrino
113
E130 Wrong sample. For SET, KFT or DOC with preset titration direction
the first measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E131 No EP set for SET, STAT.
Exit: The error message disappears on next startup.
E132 Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133 Silo full.
Exit: Send new command.
E134 No method. A method, which is required from the silo memory or in TIP, does
not exist.
Exit: The error message disappears on next startup.
E135 Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
E136 Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137 XXX Bytes are missing so that the method, the silo line could not be stored or
not enough RAM for running TIP.
Exit: Send new command.
E149 Rate missing for calculation.
Exit: The error message disappears on next start or on recalculation.
E150 Number of rates does not agree with the set windows or the mean rate C80
could not be calculated.
Exit: The error message disappears on next start or on recalculation.
E151 Fix volume missing for calculation.
Exit: The error message disappears on next startup or on recalculation.
E155 No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E157 No sequence defined in TIP.
Exit. The error message disappears on next start.
E158 A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
E160 No new temporary variable.
Exit: The error message disappears on next start.
E161 Measurement range of the secondary measured value (temperature) exceeded.
The primary measured value (pH, U, Ipol, Upol) can also be unstable.
Exit: Rectify error or &m $S.
E162 No ramp end set in DOC.
Exit: The error message disappears on next start or &m $S.
E166 Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
E167 Rate too high in DOS. No dispensing possible with the Exchange Unit currently
mounted.
Exit: The error message disappears on next start or &m $S.
E170 Fix time missing for calculation.
Exit: The error message disappears on next start or on recalculation.