Manual

4.2 Maintenance and servicing
750 Autosampler
41
4.2 Maintenance and servicing
4.2.1 General information
The 750 Autosampler was designed for years of maintenance-free use.
The X,Y,Z sample tray assembly is equipped with bushings and bear-
ings made from materials chosen to minimize the need for lubrication.
The needle drive carriage position is recalibrated at powerup and with
the start of each method. Therefore, there are no user serviceable parts
in the X,Y,Z sample tray assembly or the needle drive carriage with the
exception of changing the transfer needle (see section 4.2.2).
4.2.2 Transfer needle replacement
If the transfer needle is blocked it must be replaced. Proceed as follows
(see Fig. 11):
1
Prepare Autosampler
Stop any method which may be running by pressing
<RUN/STOP> and <ENTER>.
Park the transfer needle 55 by pressing <SYSTEM> and "Park
Needle".
Switch off instrument by means of mains switch 1313.
2
Remove old needle
Hold connection screw 2121 with open-end spanner and un-
screw needle holder 1717 together with transfer needle 55 with
open-end
7
/32" spanner (6.2621.020).
Use two open-end spanners to unscrew nut 1919 from connec-
tion screw 2121 and remove connection screw 2121 from holder
2020.
Remove old transfer needle 55 from needle holder 1717.
3
Insert new needle
Insert 6.2620.130 spare needle with already mounted ferrule
1616 in needle holder 1717.
Insert connection screw 2121 into holder 2020 from below.
Use an open-end spanner to screw nut 1919 firmly onto con-
nection screw 2121.
Screw needle holder 1717 onto connection screw 2121 by hand.
Hold connecting screw 2121 with open-end
7
/16" spanner,
tighten up needle holder 1717 with open-end
7
/32" spanner
(6.2621.020) by about half a turn.