Manual

2 Installation
750 Autosampler
18
5
Start injection
Press the <ENTER> key. The 750 Autosampler will make an
injection from the vial at position 1. Afterwards the following
display will appear:
MANUAL INJECTION
COMPLETED
6
Select "SYSTEM"
Press the <SYSTEM> key (<FUNCTION> + <4>). The fol-
lowing screen will appear:
SYSTEM SETTINGS
àà
Park Needle
Contact Closures
Loop Size ââ
7
Park needle
Press the <ENTER> key. The transfer needle will be withdrawn
from the sample vial and moved back to its parking position in
the back left-hand corner. The following display then reappears:
READY
SELECT A FUNCTION
If any difficulties should occur during the initial checkout then proceed
as described in chap. 4.3.