Owner manual

3.13 Connection of devices via remote interface
746 VA Trace Analyzer / 747 VA Stand
3-57
0
I
Control Lines
Printer
External Bus
Dos. 1
Dos. 2
Dos. 3
Dos. 4
RS 232 Interface 1
RS 232 Interface 2
Reset
2A(TH)
115V
230V
230V
746
Fig. 25: Connection of external devices to the remote interface
746 VA Trace Analyzer external
Inputs
+5V
Pin Func-
tion
21 input 1
9 input 2
22 input 3
10 input 4
23 input 5
11 input 6
24 input 7
12 input 8
Voltage Scan
at input status
0 V 1 (on)
5 V 0 (off)
Outputs
Pin Func-
tion
5 output 1
18 output 2
4 output 3
17 output 4
3 output 5
16 output 6
1 output 7
2 output 8
External device
V
CEO
40 V
IC 20 mA
Voltage Scan
at input status
0 V 1 (on)
open 0 (off)
Voltages
+5V
0V
Pin Function
15 5 V: inactive, High (H)
14 0 V: active, Low (L)
25 0 V: active, Low (L)
I 40 mA
R
i
12
Contact arrangement at socket
"Control Lines" (female)
13 1
25 14
Connecting cables to external devices must have an
appropriate 25-pin connector (male). Metrohm offers
connector sockets (K.210.9060) and the associated
housing (K.210.0002) as an option for the prepara-
tion of such cables.
No liability whatsoever will be accepted for damage arising from improper connection of devices.
Remote cable
External device