Owner manual

3.12 Connection of devices via RS232 interfaces
746 VA Trace Analyzer / 747 VA Stand
3-55
3.12.3 Connection of a balance
The balances listed below can be attached to RS232 interface 2 of the 746 VA
Trace Analyzer. The settings of the RS232 parameters on the 746 VA Trace Ana-
lyzer depend on the settings of the balance, Receive mode must be set to on.
Balance Cable
Settings on 746 VA Trace Analyzer
Sartorius MP8, MC1 6.2125.070 Balance type Sartorius
Mettler AB, AG (LC-RS25) cable from Mettler Balance type Mettler
Mettler AM, PM ME 33995 from Mettler:
green lead to pin 2,
brown to 3, white to 7,
yellow to 20 of the 25-
pin connector
Balance type Mettler
Mettler with interface 016 cable from Mettler:
red lead to pin 3, white
to 7 of the 25-pin
connector
Balance type Mettler
Mettler with interface 011
or 012
6.2125.020 Balance type Mettler
Mettler AT ME 33995 from Mettler:
green lead to pin 2,
brown to 3, white to 7,
yellow to 20 of the 25-
pin connector
Balance type Mettler AT
Precisa
Balances with RS232
interface
6.2125.080
Balance type Precisa
AND
ER-60, 120, 180, 182
FR-200, 300
FX-200, 300, 320 with
RS232 interface (OP-03)
6.2125.020
6.2125.020
6.2125.020
Balance type AND
Balance type AND
Balance type AND
A balance and a printer can be attached to RS232 interface 2 at the same time with
the aid of the 6.2125.030 Socket-outlet adapter. The balance must then be plugged
into the ”data in” port of the socket-outlet adapter.
The initial mass is transferred as a number with up to 6 digits, sign and decimal
point. Units and control characters sent by the balance are not transferred. With the
aid of a special input device supplied by the balance manufacturer, not only the ini-
tial mass but also sample identifications and methods can be put in from the bal-
ance. The addresses of the identification or method must be preselected on the in-
put device:
Balance Method Id#1 Id#2 Id#3
Sartorius METH or 27 ID.1 or 26 ID.2 or 24 C-20 or 23
Mettler (AT) D (Mthd) C (ID#1) B (ID#2) A (c20)