Owner manual

3.10 Connection of 700 Dosinos
746 VA Trace Analyzer / 747 VA Stand
3-47
7
Connect siphoning Dosino to 747 VA Stand
Using either a 6.1805.120 FEP Tubing (L = 1 m) or two 6.1805.120 FEP
Tubings and a 6.1808.000 Connecting sleeve, connect connection 1 of
the 6.1567.250 Dosing unit of the siphoning Dosino with coupling 3131
mounted on the 747 VA Stand (see Fig. 6).
Attach siphoning Dosino using the cable connected to the instrument to
socket 1818 (Dos. 4) of the 746 VA Trace Analyzer (see Fig. 2 or Fig. 4).
8
Initialize Dosinos
Switch on 746 VA Trace Analyzer using the mains switch 1111. The two
Dosing units of the rinsing and siphoning Dosino and the tubing connec-
tions are filled automatically at system start (the initialization procedure
can take several minutes!).
6.1805.100 6.1805.090
6.1805.180
6.1819.010
6.1819.010
6.1819.010
6.1819.010
6.1618.050
1
1
747
6.1805.120
6.1805.120
6.1829.020
6.1621.000
6.1805.120
6.1808.000
6.1618.050
6.1621.000
6.1808.000
6.1829.020
6.1805.120
6.1456.210
6.1808.000
Fig. 22: Scheme of the tubing connections for Rinsing and Siphoning Dosino
700 Rinsing
Dosino (Dos.3)
700 Siphoning
Dosino (Dos.4)
Waste
Rinsing
solution