3.9 Connection of 685 Dosimats
746 VA Trace Analyzer / 747 VA Stand
3-35
3.9.3 Tubing connection for addition of auxiliary solutions
The 4-way microtip 4747 is not suited for the addition of auxiliary solutions in the mL-
range into the measuring vessel of the 747 VA Stand. For the addition of a first aux-
iliary solution we recommend the use of the following accessories:
1 ×× 6.1819.000 PTFE tube, length L = 77 mm, i.d. = 0.8 mm
2 ×× 6.1805.100 FEP tubing, length L = 40 cm, i.d. = 2 mm
1 ×× 6.1808.000 Coupling
To install these accessories and ready the 685 Dosimat for automatic dispensing,
proceed as follows:
1
Mount Exchange unit on 685 Dosimat
Procedure, see section 2 of 685 Instructions for Use.
2
Insert PTFE tube in 747 VA Stand
Cut 6.1819.000 PTFE tube (optionally available) to a length of ca. 30 mm
(the tip of the tube should never be immersed into the solution).
Remove dummy stopper 4242 and insert PTFE tube from above in the
opening 7878 of the measuring head 4141 (see Fig. 9).
3
Connect FEP tubing to Exchange unit
Screw 6.1805.100 FEP tubing (optionally available) into opening 7878.
Screw a 1808.000 Coupling (accessory of 747 VA Stand) to the other end
of the 6.1805.100 FEP tubing and insert this in a slot of the tubing holder
at the rear of the 747 VA Stand.
Unscrew the 6.1543.050 Burette tip from the 6.1805.100 FEP tubing
attached to the connection 2 of the Exchange unit (see section 3.9.2).
Screw the other end of the 6.1805.100 FEP tubing attached to the Ex-
change unit to the 6.1808.000 Coupling just inserted in the tubing holder
of the 747 VA Stand.
4
Initialize Dosimat(s)
Switch on 746 VA Trace Analyzer using the mains switch 1111.
Select dialog page "DOSIMATS".
Set parameter V.init to 0.00 mL.
Set parameter V.tube in to 1.7 mL.
Set parameter V.tube out to 3.0 mL.
Press softkey <Prime all> for initializing all attached Dosimats or select
desired Dosimat with the cursor keys and press softkey <Prime Dosimat>:
The burette cylinder is filled and excess solution is expelled to the meas-
uring vessel. With standard accessories, all tubings to the measuring
vessel will have been filled at the end of the initializing procedure.
Start the initializing procedure a second time by pressing the softkey
<Prime all> or <Prime Dosimat> again.
Start the initializing procedure a third time by pressing the softkey <Prime
all> or <Prime Dosimat> again.