Owner manual

8 Technical data
746 VA Trace Analyzer / 747 VA Stand
8-4
Remote interface
Connector D-Sub 25-pin
Control lines 8 input lines, 8 output lines
Detailed description see section 3.13
Connection of Dosimats/Dosinos
Type 685 Dosimat or 700 Dosino
Number 1...4
Manual operation Dispensing, filling, adjustment of feed and filling rate
Mains connection
Voltage 115 V: 100...120 V ± 10 %
230 V: 220...240 V ± 10 %
Switching with mains voltage selector
(see section 3.2.1)
Frequency 50...60 Hz
Power consumption max. 160 VA
Fuse 2 ATH, slow-blow, high breaking capacity
Ø 5 mm, length 20 mm
Safety specifications
Construction/testing According to IEC 1010 / EN 61010 / UL 3101-1,
protection class I, degree of protection IP20
Safety directions The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference Standards met:
EN 55011 (class B), EN 55022 (class B),
EN 50081-1
Immunity to interference Standards met:
EN 50082-2, EN50093, IEC801-2, IEC801-3,
IEC801-4, IEC801-5, IEC801-6
Ambient temperature
Nominal operating range 0+45 °C
at 2080 % atmospheric humidity
Storage, transport –25+60 °C
at T 40°C: max. 95% atmospheric humidity
at T = 45°C: max. 70% atmospheric humidity
at T = 60°C: max. 35% atmospheric humidity