Owner manual

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-40
7.4.4 Problems with RDE
There are various different types of RDE:
GC: Glassy carbon electrode, can be polished with aluminium oxide, glossy
surface.
Au: Gold electrode, can be polished with aluminium oxide, mat surface.
UT: Ultra-trace graphite electrode, can be scraped and polished with alumin-
ium oxide, mat surface.
Ag: Silver electrode, can be polished with aluminium oxide, mat surface.
Pt: Platinum electrode, looks very similar to Ag, mat surface.
Low background current or unstable base line
The electrode surface must be repolished.
Has the correct RDE been used?
Exchange RDE.
Check concentration of electrolyte and pH of the solution.
Has the electrode been conditioned (e.g. with CONDC) ?
Check initial and final potential of the analysis.
If the ion concentration in the solution is too high: dilute the electrolyte.
If the concentration to be determined is considerably higher than expected:
reduce sample volume.
The background current is normally higher if RDE is used in place of MME.
A background current of several 100 nA is entirely possible.
Electrolyte solution too old: make up a new one. Its working life with organic
additives may be as short as 1 day or less.
Standard addition curves are not reproducible
Check program in the instrument (stirring time, etc.).
Check RDE (see section 3.5).
The pipetting process was not correct: was the pipetting unit used properly?
Repeat the analysis again or try out automatic standard addition with Dosi-
mat, Dosino or 6.5611.010 Pipetting equipment.
Pipetting the standard solutions must be carried out by one and the same
person or with the same instrument or the same pipette.
Check and test pipetting process. When were the pipettes last calibrated
(GLP)?
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
Are the calibration solutions too old?
Would a calibration curve be more suitable?
Peak displacement
Check and adjust pH of the solution.
Check electrolyte composition and correct if necessary.
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.