Owner manual

7.4 Troubleshooting
746 VA Trace Analyzer / 747 VA Stand
7-37
No addition
Has the correct standard solution been used or is the concentration of the
solution too low: increase the volume of the standard addition or use a higher
concentration or reduce the sample amount accordingly.
If organic components are present: carry out a UV digestion or similar.
Could it be possible that the peak is not the peak which is being looked for?
Concentration of the analyte is too high: dilute.
7.4.3 Problems with HMDE
Low background current or unstable baseline
The electrode drops or the drops do not remain hanging: check MME.
Change capillary if necessary (see section 3.4.8) or replace sealing needle
(see section 3.4.9).
Has the sample been properly degassed?
Check the concentration of the electrolyte and the pH of the solution.
Check the initial potential and final potential of the analysis.
If the ion concentration in the solution is too high: dilute the electrolyte.
If the concentration to be determined is considerably higher than expected:
reduce sample volume or change the mode (e.g. from HMDE to SMDE or
DME).
Electrolyte solution too old: make up a new one. Its working life with organic
additives may be as short as 1 day or less.
Curves look like a "starry night"
Check MME. Check for break in contact between needle and capillary. If nec-
essary, clean the MME (see section 3.4.9).
If the electrode surface is overcharged: check deposition potential and time.
No drops at the capillary: change the capillary (see section 3.4.8) or replace
sealing needle (see section 3.4.9).
Have stirring and degassing been switched off during the measurement?
Check tapping mechanism on stand. If tapping strength is too weak, turn
slotted screw 7171 on the valve block during operation in an anticclockwise di-
rection until a drop falls each time the tapper is triggered.
Electrolyte solution too old: make up a new one. Its working life with organic
additives may be as short as 1 day or less.