Owner manual

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-36
Peak displacement
Check and adjust the pH of the solution.
Check electrolyte composition and correct if necessary. Use a buffer solution
instead of an acid.
Carry out a standard addition to check whether the correct peak has been
evaluated.
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
Enter a new half-wave potential in the instrument and recalculate the results.
Check reference electrode (see section 3.6).
No peak found
The peak is only displaced: adjust the half-wave potential and recalculate the
results.
The sample concentration is too low: increase the sample volume or the
amount of sample.
The concentration of the ion to be determined is too low: use HMDE (inverse
voltammetry) instead of DME or SMDE.
Are the initial and final potentials correct?
Peak is in the highest µA range
The concentration of the ion to be determined is too high: reduce the sample
volume and carry out the analysis again.
The preconcentration time under MEAS is too high, reduce it and try the
analysis again.
Double peak
Check MME. If necessary, change capillary (see section 3.4.8) or replace
sealing needle (see section 3.4.9).
Organic components interfere with the analysis: carry out a UV digestion or
other suitable sample preparation.
If a second element is present at the same potential: add this element to the
sample and carry out the analysis again. If the second peak has become
higher then the second element is present.
Has any substance been precipitated out in the measuring vessel (e.g. lead
perchlorate standard with KCl as electrolyte)?
Try out eluents with different compositions (addition of complex formers).
Check analysis parameters.
Try another measurement mode like AC1. If one substance is reversible and
the second one irreversible, only the reversible substance is detected by
AC1.
Standard addition peaks displaced
Standard solutions have been made too acidic.
Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-
ume.