Owner manual

7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-27
Error message Meaning/Causes Rectification Exit
SUBSTANCES - ERROR 3:
Param. Mass conc.
missing
Missing value for the Mass
conc. parameter
Enter Mass conc. parameter on the
”SUBSTANCES” page
Esc
SUBSTANCES - ERROR 4:
Param. Scope f./r.half
only with linear
The f.half/r.half pa-
rameter is admissible only with
a linear baseline
Change Scope or Type parameter on
the ”SUBSTANCES” page
Esc
SUBSTANCES - ERROR 5:
Param. Scope
f./r.double missing
counterpart
Missing counterpart for the
evaluation with f.double or
r.double; in double peak
evaluation the first peak must
be evaluated with f.double
and the second peak with
r.double
Check Scope parameter for both
substances and change
Esc
SUBSTANCES - ERROR 6:
Param. Scope
f./r.double direction
conflict
Wrong assignment of the two
peaks in the double peak
evaluation; the first peak (in
sweep direction) must be
evaluated with f.double, the
second with r.double
Check Scope parameter for both
substances and change
Esc
SUBSTANCES - ERROR 7:
Undefined error
Non-specified error on the
”SUBSTANCES” page
Check the ”SUBSTANCES” page Esc
VA STAND - ERROR 1:
Missing device
Connection to 747 VA Stand
broken:
Connecting cable not
plugged in
Wrong address
Hardware fault at VA Stand
Check connection to 747 VA Stand:
Plug in connecting cable
Correct address and restart 746 VA
Trace Analyzer
Run diagnostic routine for VA Stand
E/S
VA STAND - ERROR 2:
Working electrode
faulty
Fault at MME working electr.:
No connection
Too less mercury
Bad mercury quality
Dirty or clogged capillary
Dirty or damaged sealing
needle
No inert gas pressure
Dirty electrode body
Fault at reference electrode
No solution in measuring
vessel
No electrolyte in measuring
vessel
Tapping strength in DME
operation too weak
Fault at RDE working electrode:
No connection
Electrode tip not firmly
screwed on to the drive shaft
Contaminated electrode
surface
Check MME:
Connect reference electrode with WE
cable
Fill MME with mercury (see section
3.4.2 and 3.4.7)
Clean MME (see section 3.4.9) and
use ultrapure mercury (w 0.99999)
or filtered mercury
Change capillary (see section 3.4.8)
Clean MME (see section 3.4.9) and
replace sealing needle if need be
Check inert gas connection (see
section 3.3.3)
Rub the MME body dry, clean with
distilled water of dilute acid and dry
with a lint-free cloth
see VA STAND - ERROR 4
Add measuring solution
Add a few drops of KCl 3 mol/L to the
measuring solution
Turn slotted screw 7171 on the 747 VA
Stand during operation in an anti-
clockwise direction until a drop falls
each time the tapper is triggered
Check RDE:
Connect reference electrode with WE
cable
Install RDE properly (see section 3.5)
Regenerate RDE (see section 3.5.2)
r/a