7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-23
Error message Meaning/Causes Rectification Exit
SAMPLER - ERROR 1:
Missing device
Connection to 747 VA Stand or
695 Autosampler broken:
connecting cable not
plugged in
wrong address
hardware fault in VA Stand
Check connection to 747 VA Stand or
695 Autosampler:
plug in connecting cable
restart 746 VA Trace Analyzer
run diagnostic routine for VA Stand
E/S
SAMPLER - ERROR 2:
Hardware defect U1 or
U5 EBUS
EBUS interface faulty Inform Metrohm service E/C
SAMPLER - ERROR 3:
Transfer needle blocked
Transfer needle of 695
Autosampler blocked
695 Autosampler switched
off
Wrong response of RS232
interface of 695 Autosampler
Clean transfer needle
Switch on 695 Autosampler
Inform Metrohm service
E/S
SAMPLER - ERROR 4:
No reply from RS232
No response of RS232 inter-
face of 695 Autosampler
Inform Metrohm service E/S
SAMPLER - ERROR 5:
RS232 command too long
Faulty RS232 transfer Inform Metrohm service E/S
SAMPLER - ERROR 6:
Command to RS232 not
sent
Faulty RS232 transfer Inform Metrohm service E/S
SAMPLER - ERROR 7:
RS232 communication
error
Faulty RS232 transfer Inform Metrohm service E/S
SAMPLER - ERROR 8:
Hardware defect U4 VART
Hardware fault in 695 Auto-
sampler
Inform Metrohm service E/C
SAMPLER - ERROR 9:
No small syringe
detected
No electrical connection to
small syringe of 695 Autosam-
pler
Inform Metrohm service E/C
SAMPLER - ERROR 10:
No large syringe
detected
No electrical connection to
large syringe of 695 Autosam-
pler
Inform Metrohm service E/C
SAMPLER - ERROR 11:
No rack detected
No rack mounted on 695
Autosampler
No rack type defined for
mounted rack
Mount rack
Define rack type on the ”STAND/
Rack definition” page and store data
in 695 Autosampler
E/C
SAMPLER - ERROR 12:
Hardware defect U3
EEPROM
Hardware fault Inform Metrohm service E/C
SAMPLER - ERROR 13:
Hardware defect U9
D-Latch
Hardware fault Inform Metrohm service E/C
SAMPLER - ERROR 14:
Illegal position number
Inadmissible position number
for the mounted rack
Change position number Pos. on the
”SAMPLES” and ”MONITORING” pages
E/S
SAMPLER - ERROR 15:
Illegal rack number
The specified rack type number
is not admissible
Define rack type on the ”STAND/Rack
definition” page and store data in 695
Autosampler
E/S