7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-17
Error message Meaning/Causes Rectification Exit
MEASURE - ERROR 5:
Illegal instruction for
VA Stand
The program contains inadmis-
sible instructions for operation
with the VA Stand (autosampler
instructions) on the
”OP.SEQUENCE” or
”SEGMENTS” page
Check programs on both pages and
correct
E/N
MEASURE - ERROR 6:
Missing A-formula
The called-up alternative
formula A## is not available on
the ”CALCULATION” page
Enter alternative formula A## on the
”CALCULATION” page
E/C
MEASURE - ERROR 7:
Variation parameter not
found
The parameter specified in the
VARNI instruction was not
found
Check program on the ”OP.SE-
QUENCE” or ”SEGMENTS” page and
correct
E/S
MEASURE - ERROR 8:
Backgnd subtraction
impossible
No background compensation
possible with the measurement
modes CYCMODE and
PSAMODE
Delete BACKGND instructions or
change measurement mode
E/C
MEASURE - ERROR 9:
Missing substance
The substance name specified
in the PDOS, PADD and CASE
instructions was not found
Define substance or change substance
name
E/C
MEASURE - ERROR 10:
Missing sweep block
No sweep block associated
with the substance specified in
the PDOS, PADD and CASE
instructions was found
− Call of a substance which is
not linked with a segment
− The segment is not called up
on the ”OP.SEQUENCE”
page
− Link substance to segment (<Link
to segm.>)
− Insert SEGMENT instruction with
segment name on the
”OP.SEQUENCE” page
E/C
MEASURE - ERROR 11:
Missing result
No result (EV/MC) associated
with the substance specified in
the PDOS, PADD and CASE
instructions was found
Check peak recognition or peak
evaluation for the substance in question
E/C
MEASURE - ERROR 12:
Missing substance for
calibration
− The CSOLN instruction has
found no associated sub-
stance
− The calibration mode of the
substance is not set to
cal.crv. or int.std.
− Link substance to segment (<Link
to segm.>)
− Set calibration mode for substance to
cal.crv. or int.std.
E/S
MEASURE - ERROR 13:
Missing standard
addition solution
− No solution for ADD instruc-
tion specified
− The standard addition
solution specified in the ADD
instructions was not found on
the ”SUBSTANCES/Calibra-
tion” page
− Enter solution name Soln.name in
ADD instruction
− Enter name of standard addition
solution on the ”SUBSTANCES/Cali-
bration” page
E/S
MEASURE - ERROR 14:
Mass conc. out of
working range
The mass concentration of a
calibration solution CSoln.
lies outside the working range
defined by Range min/Range
max
Change mass concentration Mass
conc. of the calibration solution or
working range (Range min/Range max)
on the ”SUBSTANCES/Calibration”
page
E/C
MEASURE - ERROR 15:
Illegal * in sample
size/S0
Entry of the asterisk in Sample
size/S0 is not admissible
when metered volumes or
concentrations have to be
calculated
Enter volume or mass in Sample
size/S0
E/N
MEASURE - ERROR 16:
Missing segment for
calibration
The CSOLN instruction has not
found an associated segment
Link substance to segment (<Link to
segm.>)
E/S