Owner manual
7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-15
Error message Meaning/Causes Rectification Exit
EVAL - ERROR 7:
AC2: illegal baseline
scope
With measurement mode
AC2MODE, only whole is
allowed for Scope
Set Scope to whole E/C
EVAL - ERROR 8:
AC2: illegal baseline
type
With measurement mode
AC2MODE, selection of polynom
or expon. for the baseline
type is not admissible
Set baseline type to AC2 rel., AC2
abs. or linear (identical to AC2
abs.)
E/C
EVAL - ERROR 9:
Wave: illegal baseline
scope
In the measurement mode
DCTMODE, only whole is
admissible for Scope
Set Scope to whole E/W
GRAPHICS - ERROR 1:
Display sector too
small
The selected display range is
too small
Expand display range (change dis-
play/plot parameters)
E/C
KEYBOARD - ERROR 1:
Keyboard hardware
defect
Keypad hardware faulty − Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 2:
HSA is ON, RS module
has not set HSA off
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 3:
HSB is ON before data
is set
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 4:
HSA and HSB are ON
before data is set
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 5:
CNT is ON in spite of
write cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 6:
CNT is not ON in spite
of read cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 7:
HSB is not set to ON at
read cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 8:
HSA is not set to OFF
at read cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 9:
HSB is not set to OFF
at read cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 10:
HSA + HSB are not set
to OFF at read cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 11:
Keyboard interface,
data not equal
Initializing error − Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 12:
EOS is not set to ON
after configuration
Initializing error − Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 13:
HSA is not set to OFF
after write cycle
Handshake error in the RS232
module
− Press Reset key
− If error reappears, inform Metrohm
service
S
KEYBOARD - ERROR 14:
Unknown keyboard error
Non-specified keypad error − Press Reset key
− If error reappears, inform Metrohm
service
S