Error message Meaning/Causes Rectification Exit
DIALOGUE - ERROR 6:
Abort <a> resets VA
Trace Analyzer
Warning that pressing the <a>
(abort) key initiates an auto-
matic reset at the 746 VA Trace
Analyzer
− press <a> to initiate the automatic
reset
− press <r> to prevent the automatic
reset
r/a
DOSIMAT - ERROR 1:
Cylinder empty
Dosimat/Dosino cylinder is
empty, no dispensing is
possible in manual operation
Fill Dosimat/Dosino cylinder by pressing
the softkey <Fill Dosimat>
E/S
DOSIMAT - ERROR 2:
Illegal Dosimat number
Wrong Dosimat number Set Dosimat number to 1...4 E/S
DOSIMAT - ERROR 3:
V.meas max too low for
dosing
Dispensing is not possible as
the inputted maximum meas-
uring vessel volume V.meas
max would be exceeded
Check dispensing instructions on the
”OP.SEQUENCE” and ”SEGMENTS”
pages or, if possible, increase V.meas
max
E/S
DOSIMAT - ERROR 4:
Dosimat overload
Dosimat/Dosino can not
dispense (drive jammed)
Press <Exchange position> key on
the ”DOSIMATS” page
E/C
DOSIMAT - ERROR 5:
Unlawful stopcock
position
Wrong stopcock position of
Dosimat/Dosino
Press <Exchange position> key on
the ”DOSIMATS” page
E/S
DOSIMAT - ERROR 6:
Dispensing volume too
large
Volume to be dispensed too
large for mounted Exchange or
Dosing Unit
Use larger Exchange or Dosing Unit or
reduce volume to be dispensed
E/S
DOSIMAT - ERROR 7:
Safety volume reached
Dispensing not possible as
safety volume of Dosimat/
Dosino would be exceeded
Increase safety volume or use larger
Exchange or Dosing Unit
E/S
DOSIMAT - ERROR 8:
Missing device
Dosimat/Dosino not attached
or connecting cable faulty
Attach Dosimat/Dosino, check con-
necting cable
E/S
DOSIMAT - ERROR 9:
No Exchange Unit
No exchange Unit mounted on
the Dosimat
Mount Exchange Unit E/S
DOSIMAT - ERROR 10:
Illegal dosing units
Inadmissible dispensing units
are mounted on the Dosinos
Mount different Dispensing Units E/S
DOSIMAT - ERROR 11:
Illegal Dilution
settings
Inadmissible dilution settings
on the "DOSIMATS" page (e.g.
V.tube out too low for
pipetting volume, cylinder
volume too low, pipetting
volume < 10 µL)
Change parameters for Dil.settings
on the "DOSIMATS" page
E/S
EVAL - ERROR 1:
Evaluation occupied
The evaluation routine can not
be used at present
Wait and retry E/C
EVAL - ERROR 2:
Evaluation not possible
No evaluation is possible with
the selected measurement
mode ”CYCMODE”
Change measurement mode E/C
EVAL - ERROR 3:
Double peak: missing
rear peak
− No peak at rear in the double
peak evaluation
− The peak at the rear has not
been evaluated
− − If there is no 2
nd
peak at rear, set
Scope to whole, f.half or r.half
− Check peak recognition for peak at
rear
E/C
EVAL - ERROR 4:
Double peak: wrong
baseline type
In the double peak evaluation,
both peaks must be evaluated
with the same baseline type
Set baseline type Type the same for
both peaks
E/C
EVAL - ERROR 5:
Double peak: missing
front peak
− No peak at front in the
double peak evaluation
− The peak at the front has not
been evaluated
− If there is no 2
nd
peak at front, set
Scope to whole, f.half or r.half
− Check peak recognition for peak at
front
E/C
EVAL - ERROR 6:
Half peak: illegal
baseline scope
With selected baseline type,
selection of f.half or
r.half is not admissible
If half-peak evaluation desired, set
baseline type to linear, otherwise set
Scope to whole or f.double or
r.double
E/C