7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-11
Error message Meaning/Causes Rectification Exit
DATA CARD - ERROR 9:
File name not found
Error in copying from or to data
card
− Attempt copying again
− If error reappears, delete file in
question
− If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 10:
File already closed
Error in copying from or to data
card
− Attempt copying again
− If error reappears, delete file in
question
− If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 11:
File already opened
Error in copying from or to data
card
− Attempt copying again
− If error reappears, delete file in
question
− If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 12:
Card is already full
No more free storage space on
data card
Use new data card or clear data E/C
DATA CARD - ERROR 13:
Invalid mode
Error in copying from or to data
card
− Attempt copying again
− If error reappears, delete file in
question
− If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 14:
File is protected
The selected file can not be
copied to the data card as a
protected file with the same
name is already present
(overwrite protection)
Disable overwrite protection with
<Secure/Unsecure> or change file
name
E/C
DATA CARD - ERROR 15:
Unexpected end of file
Error in copying from or to data
card
− Attempt copying again
− If error reappears, delete file in
question
− If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 16:
Card is write protected
The data card is write protected Disable overwrite protection or use a
different data card
E/C
DATA CARD - ERROR 17:
Undefined error
Unspecified error in data traffic
with the data card
Press Reset key E/C
DATA CARD - ERROR 18:
Bad battery charge
detected
Charge of data card battery
bad
− Copy all data on data card to instru-
ment memory
− Change battery
− Recopy data to data card
E/C
DATA PROC - ERROR 1:
Not enough measuring
points
Too few measured points
available for the smoothing
Extend measurement range or reduce
U.step
E/C
DATA PROC - ERROR 2:
Current overload, data
not evaluable
More than 2 overflow measured
values have appeared in
succession, the curve can thus
not be evaluated
Start new measurement or remove
overflow measured values manually
using the spike elimination
E/C
DATA PROC - ERROR 3:
No background sweep
found
No background sweep avail-
able
Rerecord measurement with back-
ground compensation
E/C
DIAGNOSIS - ERROR 1:
No VA Stand found
Diagnostic fault; connection to
747 VA Stand broken:
− connecting cable not
plugged in
− wrong address
Check connection to 747 VA Stand:
− plug in connecting cable
− set address correctly and restart 746
VA Trace Analyzer
E/C