7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-11
Error message Meaning/Causes Rectification Exit
DATA CARD - ERROR 9:
File name not found
Error in copying from or to data
card
Attempt copying again
If error reappears, delete file in
question
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 10:
File already closed
Error in copying from or to data
card
Attempt copying again
If error reappears, delete file in
question
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 11:
File already opened
Error in copying from or to data
card
Attempt copying again
If error reappears, delete file in
question
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 12:
Card is already full
No more free storage space on
data card
Use new data card or clear data E/C
DATA CARD - ERROR 13:
Invalid mode
Error in copying from or to data
card
Attempt copying again
If error reappears, delete file in
question
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 14:
File is protected
The selected file can not be
copied to the data card as a
protected file with the same
name is already present
(overwrite protection)
Disable overwrite protection with
<Secure/Unsecure> or change file
name
E/C
DATA CARD - ERROR 15:
Unexpected end of file
Error in copying from or to data
card
Attempt copying again
If error reappears, delete file in
question
If error reappears with this data card,
save all files and reformat data card
E/C
DATA CARD - ERROR 16:
Card is write protected
The data card is write protected Disable overwrite protection or use a
different data card
E/C
DATA CARD - ERROR 17:
Undefined error
Unspecified error in data traffic
with the data card
Press Reset key E/C
DATA CARD - ERROR 18:
Bad battery charge
detected
Charge of data card battery
bad
Copy all data on data card to instru-
ment memory
Change battery
Recopy data to data card
E/C
DATA PROC - ERROR 1:
Not enough measuring
points
Too few measured points
available for the smoothing
Extend measurement range or reduce
U.step
E/C
DATA PROC - ERROR 2:
Current overload, data
not evaluable
More than 2 overflow measured
values have appeared in
succession, the curve can thus
not be evaluated
Start new measurement or remove
overflow measured values manually
using the spike elimination
E/C
DATA PROC - ERROR 3:
No background sweep
found
No background sweep avail-
able
Rerecord measurement with back-
ground compensation
E/C
DIAGNOSIS - ERROR 1:
No VA Stand found
Diagnostic fault; connection to
747 VA Stand broken:
connecting cable not
plugged in
wrong address
Check connection to 747 VA Stand:
plug in connecting cable
set address correctly and restart 746
VA Trace Analyzer
E/C