3.4.8 Changing the capillary
Contamination of the glass capillary can necessitate its replacement (see also sec-
tions 7.4.2 and 7.4.3). In such a case, proceed as follows:
1
Remove multi-mode electrode from 747 VA Stand
Unscrew FEP tubing 5151 and 5959 from the MME, disconnect electrode
cable 3737 from MME.
Take multi-mode electrode 3838 out of measuring head 4141 while gently
tapping the MME to knock off any mercury drops on the glass capillary
into the measuring vessel.
Place multi-mode electrode 3838 in a horizontal position in the electrode
holder 114114 (see Fig. 15-3).
2
Unscrew slotted screw 9696
Using a suitable coin, unscrew slotted screw 9696 out of screw thread 9898
until the contact surface of the black O-ring at the Plexiglas wall (thin,
black stripe) is just visible below the metal thread 9898.
3
Dismantle old capillary
Position multi-mode electrode 3838 vertically in the electrode holder 114114
(see Fig. 15-1).
Undo retaining nut 108108 completely by turning anticlockwise and lift up
until the lower part of the glass capillary 109109 with the wide opening is
visible.
Gently tap the glass capillary 109109 to knock off any residual mercury in the
wide opening into the MME.
Press the retaining nut 108108 downward with one hand and with your other
hand take glass capillary 109109 completely out of the mount.
4
Dispose of old capillary
Connect filling tubing 115115 with the tubing coupling 118118 to the two gas
wash bottles 117117 and the vacuum pump (see Fig. 14).
Insert glass capillary 109109 (capillary end) in the filling cone 116116 of the filling
tubing 115115.
Remove mercury from capillary with the vacuum pump.
5
Replenish mercury if necessary
Proceed as described in section 3.4.2.
6
Mount new capillary
Proceed as described in section 3.4.3.
7
Fill capillary
Proceed as described in section 3.4.4 or section 3.4.5.