Owner manual

7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-10
Error message Meaning/Causes Rectification Exit
CURVES - ERROR 3:
No segment available
No segment data whatsoever
available
Start new measurement E
CURVES - ERROR 4:
Selected segment not
found
The selected segment (VR) was
not found
Switch to available segment E
CURVES - ERROR 5:
Can not execute,
missing data
Action can not be executed
owing to missing result data
Start new measurement E
CURVES - ERROR 6:
Can not display
calibration curve
No calibration curve available If really desired, start new measurement
with calibration
E
CURVES - ERROR 7:
Can not handle
meas.data with actual
method
The current method has been
so changed that the experi-
mental data curves can no
longer be displayed
Reload determination with original
method or start new measurement
E
CURVES - ERROR 8:
Unqualified peak
attribute for this
operation
The requested operation can
not be performed with the
selected peak (e.g. <Modify
baseline> with peak without
baseline)
Read remarks on selected peak on the
”RESULTS/Evaluation” page
E
CURVES - ERROR 9:
Can not find plot
parameters of segment
The segment curve can not be
displayed as no plot parame-
ters are available (e.g. as the
segment name has been
changed)
Reset segment name (the name may
not be changed after the determination
has been recorded)
E
DATA CARD - ERROR 1:
Hardware error
Hardware fault on data card or
unexpected write/read error
Check data card, reinsert and attempt
write/read operation again
If error reappears, repeat write/read
test with a different data card (if suc-
cessful, reformat old card)
If error reappears, inform Metrohm
service
E/C
DATA CARD - ERROR 2:
Card invalid or not
available
No data card inserted
Data card not inserted
properly
Battery of data card fully
discharged
Insert data card
Take out data card then reinsert and
push right in
Change battery of data card
E/C
DATA CARD - ERROR 3:
Battery word bad
Test pattern in RAM of data
card destroyed, i.e. data card
battery discharged
Change battery of data card E/C
DATA CARD - ERROR 4:
Low battery charge
detected
Charge of data card battery low Copy all data on data card to instru-
ment memory
Change battery
Recopy data to data card
E/C
DATA CARD - ERROR 6:
Invalid file descriptor
Data interchange with data
card not possible
Press Reset key E/C
DATA CARD - ERROR 7:
Can not open more files
simultaneously
Data interchange with data
card not possible
Press Reset key E/C
DATA CARD - ERROR 8:
File name already
exists
Error in changing file name:
Inputted new name is inadmis-
sible as a file with the same
name exists (overwrite protec-
tion)
Use a different file name or delete old
file
E/C