Owner manual

7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-9
Error message Meaning/Causes Rectification Exit
CALIBR - ERROR 6:
Missing reference
No reference to reference
substance in the calibration
with internal standard (no
evaluation possible)
No reference substance
entered
No result for reference
substance found
Enter reference substance on the
”SUBSTANCES/Calibration” page
Check peak recognition for reference
substance
E/C
CALIBR - ERROR 7:
Sample mass too low
Sample mass in measuring
vessel <0.1 pg (no evaluation
possible)
Check sample size S0 on the
”SAMPLES” page
Check program, above all the
SMPL/M and SMPL>M instructions
E/C
CALIBR - ERROR 8:
Measuring volume out of
range
Solution volume <1 mL or
> V.meas max
Check sample size S0 on the
”SAMPLES” page
Check program, above all the
SMPL/M and SMPL>M instructions
Change V.meas max
E/C
CALIBR - ERROR 9:
Missing evaluation
values
Too few measured points
(calibration solutions) for the
selected calibration mode
No calibration curve available
No results for calibration
curve found
Quantity = P.peak
inadmissible for AC2MODE
Quantity = P.peak
inadmissible for DCTMODE
with SWEEP
Increase number of measured points
(calibration solutions) or change
calibration mode
Plot new calibration curve or – if no
calibration desired – set Technique
to none
Check peak recognition for calibration
Set Quantity to I.peak
Set Quantity to I.peak or use
DCTMODE with SSWEEP
E/C
CONFIG - ERROR 1:
Bad Autosampler
configuration
Stand address and configu-
ration entry on the ”VA
TRACE ANALYZER” page do
not agree
Wrong configuration of 695
Autosampler
Change stand address to 3 or
restart VA Trace Analyzer
Install Autosampler according to
section 3.9
E/S
CONFIG - ERROR 2:
Illegal Dosimat number
Wrong Dosimat number Set Dosimat number to 1...4 E/S
CONFIG - ERROR 3:
Illegal stand number
Wrong stand number Set stand number to 1 or 2 E/S
CONFIG - ERROR 4:
Bad VA Stand
configuration
Stand address and configura-
tion entry on the ”VA TRACE
ANALYZER” page do not agree
Change stand address or restart VA
Trace Analyzer
E/C
CONFIG - ERROR 5:
Bad Dosimat
configuration
Dosimat address and configu-
ration entry on the ”VA TRACE
ANALYZER” page do not agree
Change Dosimat address or restart VA
Trace Analyzer
E/S
CONFIG - ERROR 6:
Bad Sample Changer
configuration
Address of the 717 Sample
Changer and configuration
entry on the ”VA TRACE
ANALYZER” page do not agree
Change address at 717 Sample
Changer or restart VA Trace Analyzer
E/S
CURVES - ERROR 1:
No data assigned to
substance
No result data for selected
substance available
Perform recalculation
If error reappears, perform initializa-
tion (see section 7.3.3)
If error reappears, start new meas-
urement
E
CURVES - ERROR 2:
Modify baseline with
smoothed curve only
Modify baseline is possible
only with smoothed or sub-
tracted curves
Switch to smoothed or subtracted curve
using <Select curve>
E