3.4 Multi-mode electrode (MME)
746 VA Trace Analyzer / 747 VA Stand
3-23
3.4.7 Replenishing the mercury (without changing capillary)
The multi-mode electrode 3838 can also be refilled with mercury without having to
remove the glass capillary 109109.
1
Dismantle multi-mode electrode
• Unscrew FEP tubing 5151 and 5959 from the MME. Disconnect electrode
cable 3737 from the MME.
• Take multi-mode electrode 3838 out of the measuring head 4141 and tap the
MME gently to knock off any mercury drops on the glass capillary into the
measuring vessel.
• Place multi-mode electrode 3838 horizontally in the electrode holder 114114
(see Fig. 15-3). The slotted screw 101101 is now at the top.
2
Replenish mercury
• Unscrew slotted screw 101101 using a suitable coin. If the slotted screw 101101
can not be loosened by hand, screw on locking ring 9595 and pull out of the
MME.
• Draw up mercury using the syringe 111111 with attached needle 112112 and
expel into the Hg reservoir 103103.
The Hg reservoir 103103 must never be filled more than
2
/
3
full with
mercury.
• Reinsert slotted screw 101101 into screw thread 100100 and screw flush to
surface using a suitable coin (this action may expel a few drops of mer-
cury from glass capillary 109109).
Do not turn so tightly that the cemented-in steel threaded ring 100100
becomes loose and hence jeopardizes the tightness and safety of the
MME!