3.4 Multi-mode electrode (MME)
746 VA Trace Analyzer / 747 VA Stand
3-23
3.4.7 Replenishing the mercury (without changing capillary)
The multi-mode electrode 3838 can also be refilled with mercury without having to
remove the glass capillary 109109.
1
Dismantle multi-mode electrode
Unscrew FEP tubing 5151 and 5959 from the MME. Disconnect electrode
cable 3737 from the MME.
Take multi-mode electrode 3838 out of the measuring head 4141 and tap the
MME gently to knock off any mercury drops on the glass capillary into the
measuring vessel.
Place multi-mode electrode 3838 horizontally in the electrode holder 114114
(see Fig. 15-3). The slotted screw 101101 is now at the top.
2
Replenish mercury
Unscrew slotted screw 101101 using a suitable coin. If the slotted screw 101101
can not be loosened by hand, screw on locking ring 9595 and pull out of the
MME.
Draw up mercury using the syringe 111111 with attached needle 112112 and
expel into the Hg reservoir 103103.
The Hg reservoir 103103 must never be filled more than
2
/
3
full with
mercury.
Reinsert slotted screw 101101 into screw thread 100100 and screw flush to
surface using a suitable coin (this action may expel a few drops of mer-
cury from glass capillary 109109).
Do not turn so tightly that the cemented-in steel threaded ring 100100
becomes loose and hence jeopardizes the tightness and safety of the
MME!