Owner manual

6 Data processing and evaluation
746 VA Trace Analyzer / 747 VA Stand
6-34
6
Calculation of the mass concentration ρρ
ss
The found, effective mass concentration ρ
eff,a,seff,a,s
is converted into the sought
mass concentration ρ
ss
by taking the dilution into account (see section 6.9.2):
Mass conc. = ρ
ss
= ρ
eff,a,seff,a,s
/ f
VV
7
Calculation of the total scatter MC.dev
The total scatter σ
xx
is calculated from the two error contributions of the
calibration and the measurement as described in section 6.9.3:
(σ
x x
)
22
= (σ
x,Mx,M
)
22
+ (σ
x,Cx,C
)
22
As these scatters refer to the effective mass concentration ρ
eff,a,seff,a,s
, they must
also be converted by taking the dilution into account:
MC.dev. = σ
xx
/ f
VV
Meas.dev. = σ
x,Mx,M
/ f
VV
Cal.dev. = σ
x,Cx,C
/ f
VV
To be able to calculate Cal.dev., the following conditions for the number of
measured value pairs n in the calibration must be met:
n 2 for y = bx
n 3 for y = a + bx
n 3 for y = bx + cx
4
n 4 for y = a + bx + cx
4
If this is not the case, Cal.dev.= 0 is set and the message no cal.devia-
tion put out on the "RESULTS/Evaluations" page.
The calculation of Meas.dev. uses the number of measured value pairs m in
the measurement of the sample solution. The following cases must be
distinguished:
m 4 Meas.dev. is calculated from σ
x,Mx,M
(see section 6.9.3).
m 2 If sufficient calibration points are available (see above),
Meas.dev. is estimated from Cal.dev. and the message
dev.estimated put out on the "RESULTS/Evaluations" page.
If this is not the case, Meas.dev. is calculated from σ
x,Mx,M
(see
section 6.9.3).
m = 1 If sufficient calibration points are available (see above),
Meas.dev. is estimated from Cal.dev. and the message
dev.estimated put out on the "RESULTS/Evaluations" page.
If this is not the case, Meas.dev. = 0 is set and the message
no meas.deviation put out.