Owner manual

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-311
VWASH Page: O, S Stand: A Abbreviation: VW
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. VWASH ..... V.add ........mL
Function Additional wash volume for subsequent dispensing or transfer instruction
on the 695 Autosampler.
Parameter Range; Default value / Meaning
V.add 0.010...10.000 mL; 1.000 mL
Addition volume
Additional solution volume that is delivered following the dispensing or
transfer instruction.
Description The VWASH instruction delivers an additional wash volume V.add of
hydraulic solution following a dispensing or transfer instruction on the
695 Autosampler. This ensures that the preceding solution transfer was
actually complete.
Comment The VWASH instruction refers only to the immediately following
dispensing or transfer instruction.
The VWASH instruction can be used with the following dispensing or
transfer instructions for the 695 Autosampler:
SMPL>M
DOS>M, DOS>A, DOS>F, SDOS>M, SDOS>A, SDOS>F, PDOS>M
ADD>M, PADD>M, CSOLN>M, RADD>M, RADD>F
A>M, F>A, F>M
For solution transfers on the 695 Autosampler, the following generally
holds:
Transfer volume VWASH Siphoning off Delivery
< 200 µL V.add = 0 100% 100%
V.add > 0 100% 100% + V.add
200 µL V.add = 0 105% 100% (+ 5% in waste)
V.add > 0 100% 100% + V.add
see also DOS>A, DOS/M, DOS>M, DOS>F, PDOS>M, SDOS>M, SDOS>A,
SDOS>F, ADD/M, ADD>M, PADD>M, RADD>F, RADD>M