Owner manual
5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-304
Comments • The electrode type must be defined before the TSWEEP instruction.
Only HMDE or RDE are admissible.
• The measurement mode must be defined before the TSWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
• Apart from the TSWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
• If the current measurement should be performed without measured
value recording, the TSWEEP instruction must be replaced by the
MEAS instruction. In contrast to TSWEEP, this may be in the same
segment as the measurement sweep.
• If the measurement voltage U.meas should not be switched off at the
end of the program, enter a suitable standby voltage U.standby in a
ØMEAS instruction.
• Maximum 1000 measured values can be recorded in a segment. The
number of measured values is calculated as follows:
Measurement duration t/s
Number of measured values = –––––––––––––––––––––––––
Time interval t.step
see also HMDE, RDE, MEAS