Owner manual
5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-302
TSTIR Page: O, S Stand: M, A Abbreviation: TS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. TSTIR ..... Rot.speed ......../min
Function Switch on stirrer or a preset line time.
Parameter Range; Default value / Meaning
t/s 0.1...999.9 s; -
Stirring time
The duration of stirring is entered in the line time.
Rot.speed 0, 200, 400, ..., 3000 /min; 2000 /min
Rotational speed
Setting of the number of revolutions of the stirrer per min.
Description The TSTIR instruction is used to switch on the stirrer of the 747 VA Stand
for the period preset in t/s.
Comment If the stirrer should not be switched off automatically after a definite time,
the STIR instruction can also be used.
see also STIR, ØSTIR