Owner manual

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-300
Description The SWEEP instruction starts the voltage sweep defined in the parame-
ters.
U
t
U.start
U.end
U.step
t.step
Comments The electrode type must be defined before the SWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
The measurement mode must be defined before the SWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
Apart from the SWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
With the SWEEP instruction maximum 1000 measured values can be
recorded.
see also DSWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE