Description The SWEEP instruction starts the voltage sweep defined in the parame-
ters.
U
t
U.start
U.end
U.step
t.step
Comments • The electrode type must be defined before the SWEEP instruction.
Admissible are DME, SMDE, HMDE or RDE.
• The measurement mode must be defined before the SWEEP
instruction. Admissible are DCTMODE, DPMODE, AC1MODE, AC2MODE or
SQWMODE, but not PSAMODE or CYCMODE.
• Apart from the SWEEP instruction, no other SWEEP instruction is
allowed in the same segment.
• If the final voltage U.end should not be switched off at the end of the
program, enter a suitable standby voltage U.standby in a ØMEAS
instruction.
• With the SWEEP instruction maximum 1000 measured values can be
recorded.
see also DSWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE